FDA Adverse Event Injury Summary report: N

GUNTHER TULIP FEMORAL VENA CAVA FILTER SET

MDR report key: 8534264 · Received April 20, 2019

Report

Report Number
1820334-2019-00964
Event Type
Injury
Date Received
April 20, 2019
Report Date
August 1, 2019
Manufacturer
COOK INC
Product Code
DTK
UDI-DI
10827002330167
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: INVESTIGATION ¿ INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. UNKNOWN IF THE REPORTED BLEEDING, PAIN, TIGHTNESS, SHAKINESS ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. 20 DEVICES IN LOT. NO RELEVANT NOTES ON WO FOR DEVICE (IGTCFS-65-FEM) LOT# 4020047. NO OTHER COMPLAINTS ON LOT. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: BLEEDING, PAIN, TIGHTNESS, SHAKINESS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

H10: INVESTIGATION¿ VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. UNKNOWN IF THE REPORTED BLEEDING, PAIN, TIGHTNESS, SHAKINESS ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PREVIOUS MDR WAS SUBMITTED BY WILLIAM COOK EUROPE UNDER MANUFACTURER REPORT REFERENCE# 3002808486-2018-01262. ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS INITIAL REPORT, COOK INC INFORMS THAT ALL FUTURE SUBMISSIONS REGARDING THIS COMPLAINT WILL BE HANDLED UNDER MANUFACTURER REPORT REFERENCE# (MANUFACTURER REFERENCE NUMBER). (B)(4). OCCUPATION: NON-HEALTHCARE PROFESSIONAL. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

PATIENT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2013 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO DEEP VEIN THROMBOSIS AND PULMONARY EMBOLISM. PATIENT ALLEGES VENA CAVA PERFORATION, BLEEDING. PATIENT FURTHER STATES "BLEEDING INTERNALLY AND COMING OUT EXTERNALLY, PAIN IN ABDOMEN/ PULLING, TIGHTNESS, SHAKINESS IN HANDS AND FEET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327075 GUNTHER TULIP FEMORAL VENA CAVA FILTER SET DTK COOK INC G33016 4020047 10827002330167

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening