ENDOPATH THORACIC ENDOLINEAR CUTTER W/SAFETY LOCKOUT
Report
- Report Number
- 1527736-1997-00058
- Event Type
- Malfunction
- Date Received
- January 30, 1997
- Date of Event
- January 2, 1997
- Report Date
- January 29, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- NURSE
Narratives
BROKEN FIRING TRIGGER TEETH AND BROKEN SHORT RACK TEETH.
DURING A LUNG RESECTION THE EZ45B, SERIAL NUMBER 0280, FIRED WITHOUT DIFFICULTY ON THE FIRST FIRING. AFTER IT WAS RELOADED, NO STAPLES WERE FORMED. ANOTHER EZ45B WAS OPENED, SERIAL NUMBER 0278, AND ALSO LOCKED OUT. THE CASE WAS FINISHED WITH A COMPETITIVE PRODUCT. THERE WAS NO CONSEQUENCE TO THE PATIENT. CLINICAL FOLLOW UP: 1/6/97 1350 MESSAGE AND 800 NUMBER LEFT FOR SURGEON TO CALL BACK. 1/7/97 1445 SURGEON CALLED BACK STATING HE DID NOT PERFORM THIS PROCEDURE, BUT BELIEVED A DR MAY BE THE SURGEON WHO EXPERIENCED THE DIFFICULTY. 1/7/97 1455 PAGED SALES REP WHO CALLED BACK AND ASKED HIM TO FOLLOW UP WITH THIS SURGEON REGARDING THIS INQUIRY. THE SALES REP SAID HE WOULD CALL BACK. 1/10/97 1000 REP CALLED BACK WITH SURGEON INFORMATION. 1/10/97 MESSAGE AND 800# LEFT FOR MD CALL BACK. 1/10/97 FROM THE SURGEON'S OFFICE RETURNED CALL AND STATED THE SURGEON WILL NOT BE IN THE OFFICE UNTIL THE END OF NEXT WEEK. NNCL SENT 1/22/97 SURGEON CALLED BACK AND STATED THE TWO EZ45S WOULD NOT FIRE ON FIRST FIRING. THEY WERE RELOADED BUT WOULD NOT FIRE. THE INSTRUMENTS CLOSED OK ON TISSUE, AND THE FIRING LEVER MOVED THROUGH THE FIRING MOTION, BUT NOTHING HAPPENED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH THORACIC ENDOLINEAR CUTTER W/SAFETY LOCKOUT | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | J4249K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |