HIP GENERAL DRILL BIT Ø 3.2 L 130
Report
- Report Number
- 3005180920-2019-00300
- Event Type
- Malfunction
- Date Received
- April 18, 2019
- Date of Event
- March 22, 2019
- Report Date
- July 16, 2019
- Manufacturer
- HPF SPA
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
THE REQUEST FOR MANUFACTURING PROCESS DOCUMENT REVIEW HAS BEEN SENT TO THE SUPPLIER HPF ON 18.04.2019. DOCUMENT REVIEW RECEIVED REPORTED AS FOLLOWS: ALL THE STEPS, ACCORDING TO OUR ROUTING SHEET AND RELATIVE DRAWINGS, HAVE BEEN PERFORMED CORRECTLY AS WELL AS THE DIMENSIONAL AND FUNCTIONAL CONTROLS. THERE AREN'T NONCONFORMITY ELEMENTS IN THE DOCUMENT REVIEW. THE INSTRUMENT HAS BEEN SENT TO THE SUPPLIER ON 08.05.2019. VISUAL INSPECTION PERFORMED BY R&D HIP MANAGER: THE INSTRUMENT IS BROKEN. IT IS POSSIBLE THAT THE BENDING FORCE APPLIED TO THE DRILL BIT WAS TOO HIGH PROVIDING A BREAKAGE OF THE INSTRUMENT, BUT WE DO NOT HAVE THE CERTAINTY OF THE ROOT CAUSE. VISUAL INSPECTION PREFORMED BY THE SUPPLIER HPF RECEIVED ON 19.06.2019: AS SHOWN IN THE PICTURE, THE DRILL BIT BROKE. AS CONFIRMED BY THE CUSTOMER THE DRILL BIT WAS MOVED FROM THE HOLE TOO EARLY AND OUT OF ALIGNMENT; DUE TO THE FLEXION AND THE CONSEQUENT PRESSURE THE DEVICE BROKE. THE DEVICE WAS USED NOT IN A PROPER WAY. ON JULY 16, AT THE TIME OF FU2 SENDING, WE REALIZED THAT FU1 PREPARED ON MAY 24TH FAILED DURING SENDING PROCESS. WE UPDATED FU2 WITH ALL THE INFORMATION PRESENT IN FU1.
DRILL BIT BROKE DURING THE PROCESS OF EXTRACTING THE BIT FOLLOWING DRILLING. THE SURGEON MOVED THE DRILL OUT OF ALIGNMENT TOO EARLY RESULTING IN PRESSURE ON THE BIT AND IT SNAPPING. THE INSTRUMENT WAS NO LONGER FIT FOR PURPOSE DUE TO WEAR AND TEAR. NO DEBRIS WENT INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318885 | HIP GENERAL DRILL BIT Ø 3.2 L 130 | SURGICAL INSTRUMENT FOR HIP | LXH | HPF SPA | 17H0523 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Other |