FDA Adverse Event Malfunction Summary report: N

CELL-DYN 3700 SL ANALYZER

MDR report key: 8523435 · Received April 17, 2019

Report

Report Number
2919069-2019-00035
Event Type
Malfunction
Date Received
April 17, 2019
Date of Event
March 24, 2019
Report Date
May 16, 2019
Manufacturer
ABBOTT DIAGNOSTICS DIVISION
Product Code
GKZ
UDI-DI
00380740002732
PMA / PMN Number
K980614
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE SUBSEQUENT SITE VISIT, THE FIELD SERVICE ENGINEER (FSE) RESOLVED THE ISSUE BY REPLACING THE TUBING IN THE PINCH VALVE ABOVE THE OPEN MODE PROBE (PART NUMBER 9130542). A REVIEW OF INSTRUMENT STATUS DETERMINED THAT THE RECOMMENDED PREVENTIVE MAINTENANCE HAS NOT BEEN PERFORMED ON THIS INSTRUMENT, WHICH INCLUDES REPLACEMENT OF THE TUBING AFFECTED. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. A REVIEW OF THE CELL-DYN 3700 SYSTEM OPERATOR'S MANUAL FOUND LABELING TO BE ADEQUATE FOR THE COMPLAINT ISSUE. A SEARCH FOR SIMILAR COMPLAINTS DID NOT IDENTIFY ANY ADDITIONAL COMPLAINTS OR TRENDS ASSOCIATED WITH THIS ISSUE. BASED ON THE INVESTIGATION THERE IS NO PRODUCT DEFICIENCY IDENTIFIED FOR THE CELL-DYN 3700, SERIAL NUMBER (B)(4).

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THERE IS NO FURTHER PATIENT INFORMATION PROVIDED DUE TO PRIVACY ISSUES.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FALSELY DECREASED HEMATOCRIT (HCT) RESULT ON ONE PATIENT. THE RESULTS PROVIDED WERE: (B)(6) 2019 = 23.6 / 24.5. THE CUSTOMER STATED THAT THE PATIENT WAS TRANSFUSED AS POLICY IS TO TRANSFUSE PATIENTS WITH HCTS <28. THE SAMPLE WAS REPEATED IN THE AFTERNOON AND THE HCT RESULTS WERE 30.2 / 34.3. THERE WAS NO ADVERSE IMPACT TO THE PATIENT REPORTED DUE TO THE UNNECESSARY TRANSFUSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318464 CELL-DYN 3700 SL ANALYZER AUTOMATED HEMATOLOGY ANALYZER GKZ ABBOTT DIAGNOSTICS DIVISION 00380740002732

Patients

Seq Age Sex Outcome Treatment
1 Other