TEOSYAL RHA 4
Report
- Report Number
- 3005975625-2019-00011
- Event Type
- Injury
- Date Received
- April 17, 2019
- Date of Event
- November 23, 2018
- Report Date
- April 17, 2019
- Manufacturer
- TEOXANE SA
- Product Code
- LMH
- PMA / PMN Number
- P170002
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
ACCORDING TO THE RECEIVED INFORMATION AND THE PHOTOS, THE CASE IS LIKELY TO BE A VASCULAR COMPROMISE. VASCULAR COMPLICATIONS (OCCLUSIONS) ARE WELL KNOWN AND DOCUMENTED ADVERSE REACTIONS AS PART OF HYALURONIC ACID-BASED DERMAL FILLERS INJECTIONS. THEY ARE RELATED TO THE ACCIDENTAL INJECTION OF THE PRODUCT INSIDE OR CLOSE TO A BLOOD VESSEL LEADING TO AN OCCLUSION OR A COMPRESSION AND BLOCKING THE BLOOD FLOW, GENERALLY REFERED TO AS VASCULAR COMPLICATION. IF THE VASCULAR COMPLICATION IS NOT DETECTED/DIAGNOSED AND TREATED TIMELY, IT CAN LEAD TO A SKIN NECROSIS. THE RISK OF VASCULAR COMPLICATION AND SKIN NECROSIS ARE MENTIONED IN THE PRODUCT LABELLING. BIBLIOGRAPHY: DELORENZI C. COMPLICATIONS OF INJECTABLE FILLERS, PART 2: VASCULAR COMPLICATIONS. AESTHET SURG J. 2014;34(4):584-600. FUNT D, PAVICIC T. DERMAL FILLERS IN AESTHETICS: AN OVERVIEW OF ADVERSE EVENTS AND TREATMENT APPROACHES. CLIN COSMET INVESTIG DERMATOL 2013;6:295-316. MANAFI A, BARIKBIN B, MANAFI A, HAMEDI ZS, MOGHADAM SA. NASAL ALAR NECROSIS FOLLOWING HYALURONIC ACID INJECTION INTO NASOLABIAL FOLDS: A CASE REPORT. WORLD JOURNAL OF PLASTIC SURGERY. 2015;4(1). HONG, J.Y., ET AL., IMPENDING SKIN NECROSIS AFTER DERMAL FILLER INJECTION: A "GOLDEN TIME" FOR FIRST-AID INTERVENTION. DERMATOL THER, 2017. 30(2). MARUYAMA, S., A HISTOPATHOLOGIC DIAGNOSIS OF VASCULAR OCCLUSION AFTER INJECTION OF HYALURONIC ACID FILLER: FINDINGS OF INTRAVASCULAR FOREIGN BODY AND SKIN NECROSIS. AESTHET SURG J, 2017. 37(9): P. NP102-NP108. SALVAL, A., ET AL., IMPENDING FACIAL SKIN NECROSIS AND OCULAR INVOLVEMENT AFTER DERMAL FILLER INJECTION: A CASE REPORT. AESTHETIC PLAST SURG, 2017. 41(5): P. 1198-1201. WANG, Q., ET AL., VASCULAR COMPLICATIONS AFTER CHIN AUGMENTATION USING HYALURONIC ACID. AESTHETIC PLAST SURG, 2018. 42(2): P. 553-559.
THIS EVENT HAPPENED OUTSIDE OF THE U.S., IN (B)(6). ACCORDING TO THE RECEIVED INFORMATION, TWO DAYS AFTER THE INJECTIONS, THE PATIENT EXPERIENCED A DARK BLUE AREA IN THE LEFT TEMPLE, WITH RED DISCOLORATIONS IN THE HAIRLINE, WHICH HAS BEEN CONSIDERED AS A VASCULAR COMPROMISE. AN EXPERT WAS CONSULTED BY THE INJECTOR AND A TREATMENT WITH HYALASE 3000 IU WAS INITIATED ON THE DAY OF THE REACTION, WITH ONE SESSION IN THE MORNING AND A SECOND ONE IN THE EVENING. THIS INDUCED A REVASCULARIZATION OF THE AREA AND A THIRD HYALURONIDASE SESSION WAS DEEMED NOT NECESSARY BY THE INJECTOR AND THE PATIENT. ON (B)(6) 2019, THE RESOLUTION HAS BEEN REPORTED AS COMPLETE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 317526 | TEOSYAL RHA 4 | HYALURONIC ACID DERMAL FILLER GEL | LMH | TEOXANE SA | NOT APPLICABLE | TPUL-180324A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Other |