BD VACUTAINER® EDTA 2K
Report
- Report Number
- 1917413-2019-01185
- Event Type
- Malfunction
- Date Received
- April 16, 2019
- Date of Event
- March 31, 2019
- Report Date
- April 23, 2019
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- PMA / PMN Number
- K981013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES AND PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR LEAKAGE FROM MOLDING DEFECT WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, RETENTION SAMPLES WERE INSPECTED FROM BD INVENTORY FOR EVALUATION OF MOLD B30 CAVITY 42 NONE WERE FOUND. TUBES FROM MOLD B30 CAVITY 42 WERE MOLDED AND FOUND NO MOLDING DEFECTS. A REVIEW OF THE DEVICE HISTORY RECORDS FOR FINISHED PRODUCT AND TUBE WERE COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA AND POTENTIAL CAUSES HAVE BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS HAVE BEEN ESTABLISHED AND ARE IN THE PROCESS OF BEING IMPLEMENTED. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR LEAKAGE FROM MOLDING DEFECT WITH THE INCIDENT LOT WAS OBSERVED. EVALUATION OF THE CUSTOMER SAMPLES WAS CONDUCTED, AND LEAKAGE FROM MOLDING DEFECT WAS OBSERVED. ADDITIONALLY, EVALUATION OF THE RETAIN SAMPLES AND NEWLY MOLDED PRODUCT WAS CONDUCTED. FURTHER INVESTIGATION ACTIVITIES HAVE BEEN CONDUCTED THROUGH A CAPA AND THE MOST LIKELY ROOT CAUSE HAS BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS AND PROCEDURES ARE BEING IMPLEMENTED TO MITIGATE FURTHER OCCURRENCES. ROOT CAUSE DESCRIPTION: A CAPA WAS CONDUCTED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION HAS IDENTIFIED THE MOST LIKELY ROOT CAUSES AND CORRECTIVE ACTIONS ARE IN THE PROCESS OF BEING IMPLEMENTED. CAPA 893122. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE. THE INVESTIGATION HAS IDENTIFIED POTENTIAL ROOT CAUSE(S) FOR THIS ISSUE AND CORRECTIVE ACTIONS ARE IN THE PROCESS OF BEING IMPLEMENTED.
IT WAS REPORTED THAT THE BD VACUTAINER® EDTA 2K EXPERIENCED LEAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: BLOOD LEAKED FROM THE GATE OF THE TUBE.
INITIAL REPORTER PHONE# (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD VACUTAINER® EDTA 2K EXPERIENCED LEAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: BLOOD LEAKED FROM THE GATE OF THE TUBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314447 | BD VACUTAINER® EDTA 2K | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 8249516 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |