SPRINT
Report
- Report Number
- 2649622-2019-06396
- Event Type
- Injury
- Date Received
- April 16, 2019
- Date of Event
- March 18, 2019
- Report Date
- December 27, 2024
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
CONCOMITANT MEDICAL PRODUCTS: DDMB1D1 ICD, IMPLANTED ON (B)(6) 2017. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS, HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE OF THE SUPERIOR VENA CAVA DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY ALSO INDICATED THE IMPEDANCE TREND OF THE SUPERIOR VENA CAVA DEFIBRILLATION COIL WAS VARIABLE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT AN AUDIBLE ALERT WAS TRIGGERED FOR THE RIGHT VENTRICULAR (RV) LEAD DUE TO HIGH OUT-OF-RANGE SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310855 | SPRINT | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694265 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Male | Required Intervention | 6940-52 LEAD| 6940-52 LEAD |