MILAGRO ADVANCE SCREW 7X23MM
Report
- Report Number
- 1221934-2019-56859
- Event Type
- Malfunction
- Date Received
- April 16, 2019
- Date of Event
- March 1, 2019
- Report Date
- March 20, 2019
- Manufacturer
- DEPUY MITEK LLC US
- Product Code
- MAI
- UDI-DI
- 10886705022243
- PMA / PMN Number
- K143660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY. UDI: (B)(4). INVESTIGATION SUMMARY ==> THE COMPLAINT DEVICE IS NOT BEING RETURNED, IT WAS DISCARDED BY THE CUSTOMER, THEREFORE, UNAVAILABLE FOR A PHYSICAL EVALUATION. THIS COMPLAINT CANNOT BE CONFIRMED. NR-0090833 WAS IDENTIFIED FOR THIS LOT NUMBER. THE COMPLAINT CONDITION IS UNRELATED TO THE NON-CONFORMANCE. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. IF ANY ADDITIONAL INFORMATION IS OBTAINED, THIS COMPLAINT WILL BE RE-OPENED TO CAPTURE THAT INFORMATION. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.
DATE OF EVENT IS AN UNKNOWN DATE IN (B)(6) 2019. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. (B)(4). DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED BY THE AFFILIATE THAT DURING A LIGAMENTOPLASTY BTB THE MILAGRO ADVANCE BROKE ON THE SECOND BOLT TURN. FRAGMENTS WERE GENERATED FROM THE BROKEN DEVICE AND WERE REMOVED FROM THE PATIENT. IT IS UNKNOWN IF THERE WERE ANY SURGICAL DELAY, MEDICAL INTERVENTION PLANNED, OR IF THE CASE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 313664 | MILAGRO ADVANCE SCREW 7X23MM | ORTHOPAEDIC BONE SCREW, BIOABSORBABLE | MAI | DEPUY MITEK LLC US | 231816 | L705405 | 10886705022243 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |