FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

MDR report key: 8517390 · Received April 16, 2019

Report

Report Number
8010047-2019-01692
Event Type
Malfunction
Date Received
April 16, 2019
Date of Event
March 15, 2019
Report Date
April 24, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) BUT WAS RETURNED TO OLYMPUS FRANCE (OFR). OFR SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. THE ADDITIONAL MICROBIOLOGICAL TESTING INDICATED NO MICROBIAL GROWTH FOR THE ALL CHANNEL OF THE DEVICE. THEREFORE, THE TESTING RESULT CLEARED THE FRENCH GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO OMSC FOR EVALUATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE TESTED POSITIVE FOR SERRATIA MARCESCENS (8 CFU/100ML) AND PSEUDOMONAS AERUGINOSA (> 100 CFU/100ML). THE DEVICE HAD BEEN REPROCESSED USING A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR MODEL SOLUSCOPE S4 (SOLUSCOPE) WITH PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310388 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE GASTROINTESTINAL VIDEOSCOPE FDS OLYMPUS MEDICAL SYSTEMS CORP. GIF-H185

Patients

Seq Age Sex Outcome Treatment
1