FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 8517228 · Received April 16, 2019

Report

Report Number
3008382007-2019-00919
Event Type
Malfunction
Date Received
April 16, 2019
Report Date
April 5, 2019
Manufacturer
LIFESCAN EUROPE GMBH
Product Code
NBW
UDI-DI
00353885008402
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

COMPLAINTS RELATING TO THE REPORTED PRODUCT(S) WERE EVALUATED. IT WAS CONCLUDED THAT THE NUMBER OF COMPLAINTS FOR THE PRODUCT(S) DID NOT BREACH THRESHOLDS INDICATIVE OF A SYSTEMIC ISSUE

Description of Event or Problem · 1

ON (B)(6) 2019, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER READ INACCURATELY ERRATIC. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿254, 105 AND 119 MG/DL¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR PRECISION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310773 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE GMBH 4463000 00353885008402

Patients

Seq Age Sex Outcome Treatment
1 65 YR