FDA Adverse Event Malfunction Summary report: N

TANDEM T:SLIM INSULIN DELIVERY SYSTEM

MDR report key: 8516419 · Received April 15, 2019

Report

Report Number
3013756811-2019-09740
Event Type
Malfunction
Date Received
April 15, 2019
Date of Event
January 1, 2019
Report Date
April 15, 2019
Manufacturer
TANDEM DIABETES CARE
Product Code
LZG
PMA / PMN Number
K111210
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

POSSIBLE USE ERRORS: OVERFILLING THE CARTRIDGE BAG. PER TANDEM'S T:SLIM USER GUIDE: "THE SINGLE-USE DISPOSABLE CARTRIDGE CAN HOLD UP TO 300 UNITS (3.0 ML) OF INSULIN." PUMP SERIAL NUMBERS: (B)(4). CARTRIDGE LOT NUMBER: M140162.

Description of Event or Problem · 1

QUARTERLY SUMMARY REPORT FOR ALLEGED CARTRIDGE ALARM 1: IT WAS REPORTED THAT A CARTRIDGE ALARM 1 OCCURRED DURING THE LOAD SEQUENCE OR DURING BASAL/BOLUS DELIVERY. THE ISSUE WAS RESOLVED BY LOADING A NEW CARTRIDGE AND/OR INFUSION SET, OR THE USER REVERTED TO AN ALTERNATE METHOD OF INSULIN DELIVERY. THERE WAS NO ADVERSE IMPACT TO THE USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308558 TANDEM T:SLIM INSULIN DELIVERY SYSTEM INSULIN PUMP LZG TANDEM DIABETES CARE 004628