FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 8514545 · Received April 15, 2019

Report

Report Number
1645337-2019-10261
Event Type
Injury
Date Received
April 15, 2019
Date of Event
August 29, 2018
Report Date
March 25, 2019
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001294
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 4/19/19, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 5/8/19, A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED FOR THE FINISHED DEVICE NUMBER 214041, AND NO NON-CONFORMANCES RELATED TO THE REPORTED COMPLAINT WERE IDENTIFIED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 5/20/2019, DURING ANALYSIS OF THE DEVICE IT WAS OBSERVED TO BE INTACT. SINCE THE AUTHORIZATION FOR RETURN AND EXAMINATION OF MEDICAL DEVICE FORM WAS NOT SIGNED AND/OR RETURNED, MENTOR PRODUCT ANALYSIS LAB WAS PRECLUDED FROM FURTHER EVALUATING THE DEVICE. NO OTHER ANOMALIES WERE DISCOVERED. A POSSIBLE CAUSE FOR THE CONDITION REPORTED IS DUE TO EXCESSIVE FORCE TO THE CHEST; TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. THE REPORTED COMPLAINT WAS CONFIRMED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. BASED ON THE INFORMATION REPORTED AND/OR THE PRODUCT INVESTIGATION, THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH THE MANUFACTURING PROCESS. NO FURTHER INVESTIGATION WILL BE CONDUCTED ON THIS COMPLAINT DUE THE EVENT REPORTED WAS NOT CONFIRMED. THIS REPORT IS NOT INTENDED TO DENY THAT PATIENTS EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH A SALINE MENTOR SMOOTH ROUND MODERATE PROFILE 475CC AND EXPERIENCED DEFLATION ON THE LEFT BREAST IMPLANT. AS A RESULT, THE PATIENT HAD UNDERGONE REMOVAL AND REPLACEMENT WITH GEL MENTOR MEMORYGEL BREAST IMPLANT 465CC ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308086 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 214041 00081317001294

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention