AXIOM ICONOS R200
Report
- Report Number
- 3004977335-2019-77543
- Event Type
- Malfunction
- Date Received
- April 15, 2019
- Date of Event
- April 4, 2019
- Report Date
- April 10, 2019
- Manufacturer
- SIEMENS HEALTHCARE GMBH
- Product Code
- JAA
- PMA / PMN Number
- K992660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THE REPORTED ISSUE WAS INVESTIGATED WITH THE FOLLOWING RESULT. THE DESCRIBED BEHAVIOR WAS CAUSED BY A DE-CALIBRATED JOYSTICK. THIS BEHAVIOR COULD BE REPRODUCED AS SHOWN IN THE PROVIDED VIDEOS. FOLLOWING THE RE-ADJUSTMENT OF THE JOYSTICK BY THE SERVICE TECHNICIAN, THE ISSUE AT THE CONCERNED SITE WAS RESOLVED. THE JOYSTICK WAS LATER EXCHANGED FOLLOWING THE SUCCESSFUL RE-CALIBRATION AND REQUESTED FOR FURTHER INVESTIGATION, HOWEVER, IT WAS NOT PROVIDED DESPITE SEVERAL REQUESTS. THEREFORE, A DETAILED ANALYSIS OF THE JOYSTICK WAS NOT POSSIBLE. THE SPARE PART CONSUMPTION OF THE CONCERNED PART (CROSS SWITCH WITH POTI, MATERIAL NUMBER 7030765) SHOWS VALUES BELOW THE DEFINED THRESHOLD. IN THE MAINTENANCE INSTRUCTIONS THE ADJUSTMENT OF THE JOYSTICK IS PART OF THE REQUIRED ACTIONS (SEE XPD3-340.831.20.07.02, CHAPTER 5.2.12.2 OPTIONS, CHECKING THE UNIT CONTROL CONSOLE FOR FUNCTION AND DAMAGE).
INVESTIGATION OF THE REPORTED EVENT IS ON-GOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(6).
SIEMENS BECAME AWARE OF UNINTENDED SYSTEM MOVEMENT OF THE AXIOM ICONOS R200. DURING A PROCEDURE THE OPERATOR NOTICED THAT THE TABLE SLOWLY TILTED WITHOUT ANY CONTROL KEY OR LEVER BEING TOUCHED ON THE CONTROL DESK. THE MOVEMENT WAS PERFORMED TOWARDS THE HEAD POSITION FROM 0° (TABLE HORIZONTAL) TO MINUS DEGREES. THERE ARE NO INJURIES ATTRIBUTED TO THIS EVENT. THE REPORTED EVENT OCCURRED IN (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 308427 | AXIOM ICONOS R200 | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM | JAA | SIEMENS HEALTHCARE GMBH | 10093962 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |