FDA Adverse Event Malfunction Summary report: N

AXIOM ICONOS R200

MDR report key: 8514017 · Received April 15, 2019

Report

Report Number
3004977335-2019-77543
Event Type
Malfunction
Date Received
April 15, 2019
Date of Event
April 4, 2019
Report Date
April 10, 2019
Manufacturer
SIEMENS HEALTHCARE GMBH
Product Code
JAA
PMA / PMN Number
K992660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE WAS INVESTIGATED WITH THE FOLLOWING RESULT. THE DESCRIBED BEHAVIOR WAS CAUSED BY A DE-CALIBRATED JOYSTICK. THIS BEHAVIOR COULD BE REPRODUCED AS SHOWN IN THE PROVIDED VIDEOS. FOLLOWING THE RE-ADJUSTMENT OF THE JOYSTICK BY THE SERVICE TECHNICIAN, THE ISSUE AT THE CONCERNED SITE WAS RESOLVED. THE JOYSTICK WAS LATER EXCHANGED FOLLOWING THE SUCCESSFUL RE-CALIBRATION AND REQUESTED FOR FURTHER INVESTIGATION, HOWEVER, IT WAS NOT PROVIDED DESPITE SEVERAL REQUESTS. THEREFORE, A DETAILED ANALYSIS OF THE JOYSTICK WAS NOT POSSIBLE. THE SPARE PART CONSUMPTION OF THE CONCERNED PART (CROSS SWITCH WITH POTI, MATERIAL NUMBER 7030765) SHOWS VALUES BELOW THE DEFINED THRESHOLD. IN THE MAINTENANCE INSTRUCTIONS THE ADJUSTMENT OF THE JOYSTICK IS PART OF THE REQUIRED ACTIONS (SEE XPD3-340.831.20.07.02, CHAPTER 5.2.12.2 OPTIONS, CHECKING THE UNIT CONTROL CONSOLE FOR FUNCTION AND DAMAGE).

Additional Manufacturer Narrative · 1

INVESTIGATION OF THE REPORTED EVENT IS ON-GOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(6).

Description of Event or Problem · 1

SIEMENS BECAME AWARE OF UNINTENDED SYSTEM MOVEMENT OF THE AXIOM ICONOS R200. DURING A PROCEDURE THE OPERATOR NOTICED THAT THE TABLE SLOWLY TILTED WITHOUT ANY CONTROL KEY OR LEVER BEING TOUCHED ON THE CONTROL DESK. THE MOVEMENT WAS PERFORMED TOWARDS THE HEAD POSITION FROM 0° (TABLE HORIZONTAL) TO MINUS DEGREES. THERE ARE NO INJURIES ATTRIBUTED TO THIS EVENT. THE REPORTED EVENT OCCURRED IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308427 AXIOM ICONOS R200 IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM JAA SIEMENS HEALTHCARE GMBH 10093962

Patients

Seq Age Sex Outcome Treatment
1 Other