IMPLANTIUM
Report
- Report Number
- 3005503242-2007-00006
- Event Type
- Other
- Date Received
- May 3, 2007
- Date of Event
- January 1, 2006
- Report Date
- May 23, 2006
- Manufacturer
- BIOMEDICARE, INC.
- Product Code
- DZE
- PMA / PMN Number
- K041368
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- DENTIST
Narratives
COMPLAINT WAS NOT INITIALLY INVESTIGATED. COMPLAINT PROCEDURE RECENTLY REVISED TO REQUIRE APPROPRIATE INVESTIGATION AND REPORTING OF EVENTS. REPORT BEING SUBMITTED SUBSEQUENT TO COMPLETION OF INVESTIGATION. TYPE OF TEST(S): SEND THE RETURNED PRODUCT TO THE LAB FOR SEM (SCANNING ELECTRON MICROSCOPE) IMAGING TO VERIFY SLA (SAND BLASTING WITH LARGE GRIT ACID ETCHING) TYPE SURFACE TREATMENT WAS COMPLETED. RE-INSPECT THE RETURNED PRODUCT'S DIMENSIONS TO VERIFY IF THE PRODUCT MEETS ITS SPECIFICATIONS. RESULTS: SEE ATTACHED TEST DATA. AS A RESULT OF A SEM IMAGING, SLA TYPE SURFACE TREATMENT WAS COMPLETED. AS A RESULT OF A RE-INSPECTION, THE PRODUCT MEETS ITS SPECIFICATIONS. CONCLUSION: BASED ON INSPECTION AND TEST RESULTS, THE RETURNED PRODUCT HAS BEEN FOUND TO BE WITHIN SPECIFICATIONS. THEREFORE, THE IMPLANT FAILURE IS MOST LIKELY DUE TO CAUSES OTHER THAN THE PRODUCT SUCH AS SURGICAL MISTAKE, PATIENT BONE CONDITION, PATIENT ORAL HYGIENE, OR PATIENT BEHAVIOR.
INITIAL REPORT WAS SIZE ERROR. ON INVESTIGATION, DENTIST SUBSEQUENTLY STATED THAT PATIENT'S BONE WAS TOO SOFT RESULTING TWO FAILED IMPLANTS. PATIENT RECEIVED A THIRD IMPLANT THAT WAS SUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMPLANTIUM | ENDOSSEOUS DENTAL IMPLANT | DZE | BIOMEDICARE, INC. | FX4810 | 05I21-D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Required Intervention |