BIPOLAR LEAD
Report
- Report Number
- 1644487-2007-00619
- Event Type
- Other
- Date Received
- May 14, 2007
- Date of Event
- November 1, 2004
- Report Date
- September 13, 2006
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
NO ANOMALIES WERE IDENTIFIED WITH RETURNED PORTION OF THE LEAD WHICH MAY HAVE CONTRIBUTED TO THE REPORTED HIGH LEAD IMPEDANCE. SUSPECTED USER ERROR CAUSED EVENT.
REPORTER INDICATED A VNS PATIENT UNDERWENT GENERATOR AND LEAD REPLACEMENT DUE TO LEAD DISCONTINUITY. FOLLOW-UP WITH THE RE-IMPLANTING SURGEON REVEALED THAT NO LEAD BREAK HAD OCCURRED, BUT RATHER THE "LEAD HAD PULLED OFF THE NERVE AND WAS FURTHER DOWN THE CAROTID SHEATH." PRODUCT WAS RETURNED TO MANUFACTURER AND AN "ANALYSIS WAS PERFORMED ON THE RETURNED LEAD PORTION AND THE REPORTED LEAD DISCONTINUITY CONDITION WAS NOT VERIFIED. NOTE THAT THE ELECTRODE SECTION WAS NOT RETURNED FOR ANALYSIS AND THEREFORE, A COMPLETE EVALUATION COULD NOT BE PERFORMED ON THE ENTIRE LEAD PRODUCT." A REVIEW OF THE OPERATIVE NOTES FROM THE ORIGINAL IMPLANT PROCEDURE INDICATES THAT ONLY ONE SYSTEM DIAGNOSTIC TEST WAS PERFORMED, AND THIS TEST WAS PERFORMED PRIOR TO ANCHORING THE ELECTRODES. ALL SUBSEQUENT DIAGNOSTIC TEST RESULTS IN THE PROGRAMMING HISTORY DATABASE INDICATE HIGH LEAD IMPEDANCE. USER ERROR IS SUSPECTED BECAUSE THE ORIGINAL OPERATIVE NOTES DO NOT MENTION THAT A STRAIN RELIEF BEND OR LOOP WAS CREATED PER CYBERONICS LABELING. ALSO, IT WAS NOTED THAT THE PREVIOUS IMPLANTED LEAD WAS 302-30. DURING THE REVISION SURGERY, THE SURGEON ELECTED TO IMPLANT THE SMALLER LEAD, 302-20. SINCE REPLACEMENT SURGERY "THE PATIENT HAS BEEN ABLE TO FEEL STIMULATION...BEFORE THE SURGERY THE PATIENT COULD NOT FEEL STIMULATION."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIPOLAR LEAD | LYJ | CYBERONICS, INC. | 302-30 | 010024 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Hospitalization| O| R |