FDA Adverse Event Injury Summary report: N

NAVIO

MDR report key: 8510375 · Received April 12, 2019

Report

Report Number
3010266064-2019-00002
Event Type
Injury
Date Received
April 12, 2019
Date of Event
March 15, 2019
Report Date
September 29, 2022
Manufacturer
BLUE BELT TECHNOLOGIES, DIV OF SMITH AND NEPHEW
Product Code
OLO
PMA / PMN Number
K180271
Removal / Correction Number
Z-1634-2020
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: B1, B2 AND H1 WERE UPDATED TO REPORT TYPE ADVERSE EVENT.

Additional Manufacturer Narrative · 0

THE NAVIO SOFT TISSUE PROTECTOR, (PN 101092), USED IN TREATMENT WAS NOT RETURNED FOR A PRIOR INVESTIGATION. DHR REVIEW FOUND THAT NO CONDITIONS THAT COULD CONTRIBUTE TO THE REPORTED EVENT WERE FOUND. THE REPORTED PRODUCT MET MANUFACTURING SPECIFICATIONS PRIOR TO BEING RELEASED FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTS, THIS ISSUE WILL CONTINUE TO BE MONITORED. THE SURGICAL TECHNIQUE GUIDE PROVIDES INSTRUCTIONS FOR USING THE TISSUE PROTECTOR. SPECIFICALLY, THE GUIDE PROVIDES INSTRUCTION ON HOW TO PREPARE THE BONE PIN INSERTION LOCATION ON THE PATIENT AND HOW TO INSERT THE TISSUE PROTECTOR WITHIN THAT LOCATION. THE COMPLAINT DOES NOT SUGGEST THAT THE USER DEVIATED FROM THESE INSTRUCTIONS. MOREOVER, AS PART OF THE FUNCTIONAL EVALUATION IN TR0979 THAT REPLICATED THE ISSUE, THE TEST OPERATOR FOLLOWED THE INSTRUCTIONS PROVIDED IN THE SURGICAL TECHNIQUE GUIDE AND EXPERIENCED THE BONE PIN GETTING STUCK IN THE TISSUE PROTECTOR. ACCORDINGLY, PRODUCT LABELING HAS BEEN RULED OUT AS A CAUSE OF THE COMPLAINT. THIS FAILURE IS AN IDENTIFIED FAILURE MODE WITHIN THE RISK FILE. WE COULD NOT CONFIRM IF THERE WAS A RELATIONSHIP ESTABLISHED BETWEEN THE REPORTED EVENT AND THE DEVICE. PHOTOS OF THE DEVICE WERE NOT PROVIDED FOR EVALUATION AND THE DEVICE WAS NOT RETURNED. HOWEVER, BASED ON PRIOR COMPLAINTS RECEIVED IT IS LIKELY THAT THE EVENT OCCURRED DUE TO THE REPORTED FAILURE. THE MALFUNCTION IS DUE TO A DESIGN ISSUE DUE TO THE INNER DIAMETER OF THE TISSUE PROTECTOR LUMEN DIAMETER RELATIVE TO THE MAJOR DIAMETER OF THE BONE PIN. BINDING OF THE BONE SCREW TO THE TISSUE PROTECTOR IS PRIMARILY DUE TO TISSUE BEING WRAPPED AROUND THE THREADS. HOWEVER, INITIAL PIN MISALIGNMENT AND BENDING OF THE PIN ARE ALSO CONTRIBUTING FACTORS. HHE-2020-12-PL AND CAPA 200017 WERE OPENED AS CORRECTIVE ACTIONS TO ADDRESS THIS ISSUE. AS A RESULT OF THE REMEDIAL INVESTIGATION, WE HAVE THOROUGHLY INVESTIGATED THE COMPLAINT PER THE CRITERIA AS REQUIRED BY 21 CFR 820.198(D).

Description of Event or Problem · 1

IT WAS REPORTED THAT BONE SCREW BENT WHEN DRILLING INTO FEMUR THROUGH THE TISSUE PROTECTOR. THE SURGEON WAS ABLE TO USE AN OSTEOTOME AND MALLET TO REMOVE THE TISSUE PROTECTOR AND THEN USED A DRILL TO REMOVE THE PIN. NO PATIENT INJURY OR FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300550 NAVIO STEROTAXIC INSTRUMENT, COMPUTER ASSISTED OLO BLUE BELT TECHNOLOGIES, DIV OF SMITH AND NEPHEW

Patients

Seq Age Sex Outcome Treatment
1 53 YR Female Required Intervention