STREAMLINE ISS IV POLE
Report
- Report Number
- 1825014-2019-00001
- Event Type
- Malfunction
- Date Received
- April 12, 2019
- Date of Event
- March 14, 2019
- Report Date
- April 12, 2019
- Manufacturer
- COMDEL INNOVATION
- Product Code
- FOX
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- MI, US
- Reporter Occupation
- 003
Narratives
MANUFACTURER RECORDS INVESTIGATION SHOWS THAT A DESIGN REVISION TOOK PLACE ON 4/25/14 WHERE THE SPOT TYPE WELD WAS CHANGED TO A FULL 360 DEGREE WELD AT THIS CONNECTION POINT. THE MANUFACTURER RECORDS SHOW THAT POLES SHIPPED BEFORE THIS REVISION WOULD HAVE THE SPOT WELD. THERE ARE 8 POLES THAT SHIPPED BEFORE THE REVISION AND THE SHIP TO LOCATION OF EACH HAS BEEN IDENTIFIED. TWO POLES HAVE BEEN RETURNED FROM THE HOSPITAL WHERE THE COMPLAINT ORIGINATED. THESE ARE NOW ON HOLD AT (B)(4) AND WILL BE RETURNED TO THE MANUFACTURER, COMDEL. THREE POLES ARE ON-SITE AT THE MANUFACTURER, COMDEL, AND WILL BE REWORKED WITH A 360 DEGREE WELD. THREE POLES WERE SHIPPED TO DISTRIBUTORS AS DEMO MODELS. THE MANUFACTURER WILL SHIP REPLACEMENT POLES TO THESE DISTRIBUTORS AND THE AFFECTED POLES WILL BE RETURNED TO THE MANUFACTURER TO BE REWORKED WITH A 360 DEGREE WELD.
A USER AT A HOSPITAL IDENTIFIED THAT A IV STREAMLINE POLE AND WING BARS WERE TILTED ON THE STAND. INSPECTION FOUND THE CONNECTION POINT BETWEEN THE BASE AND THE TREE SECTION WAS LOOSE. FURTHER INSPECTION FOUND THAT THE 4 VERY SMALL SPOT WELDS HAD BROKE LOOSE WHICH IS WHY IT CREATED A LOOSE FITTING FOR THE IV WING BAR SYSTEM. IV POLE WAS TAKEN OUT OF SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300510 | STREAMLINE ISS IV POLE | IV POLE | FOX | COMDEL INNOVATION | ISS-NW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |