Q6 ASPIRATION CATHETER
Report
- Report Number
- 3011410509-2019-00001
- Event Type
- Injury
- Date Received
- April 11, 2019
- Date of Event
- February 20, 2019
- Report Date
- April 11, 2019
- Manufacturer
- MIVI NEUROSCIENCE, INC.
- Product Code
- DQY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE WAS RECEIVED IN TWO PIECES. UPON VISUAL INSPECTION, DEVICE WAS SEPARATED AT THE PROXIMAL EDGE OF THE CATHETER AT THE ATTACHMENT BETWEEN THE CATHETER AND THE PUSH WIRE. THE PROXIMAL EDGE OF THE CATHETER DISPLAYED PHYSICAL DAMAGE WHERE THE PUSH WIRE WAS ATTACHED. THIS TYPE OF DAMAGE CAN OCCUR AS A RESULT OF FORCEFULLY RETRACTING THE Q CATHETER AGAINST RESISTANCE. IT IS POSSIBLE THE Q6 CATHETER INTERACTED WITH THE SHEATH WHILE ATTEMPTING TO FORCIBLY RETRACT IT AND THE DEVICE SEPARATED. ACCESSORY DEVICES WERE NOT RETURNED FOR ANALYSIS. MANUFACTURING RECORDS FOR THE DEVICE LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY RECORDS. THIS DEVICE IS A CE MARKED DEVICE DISTRIBUTED ONLY IN THE EU.
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING THE Q6 ASPIRATION CATHETER (Q6-69145-E). DURING THE PROCEDURE THE PHYSICIAN NOTICED THE PROXIMAL MARKER BAND OF THE Q CATHETER HAD BEEN EXTENDED PAST THE DISTAL MARKER BAND OF THE GUIDE CATHETER. THE PHYSICIAN ATTEMPTED TO RETRACT THE Q6 INTO THE GUIDE CATHETER BUT WAS UNSUCCESSFUL. THE PHYSICIAN THEN ATTEMPTED TO REMOVE THE Q6 AND THE CATHETER TOGETHER THROUGH THE SHEATH. DURING THE REMOVAL, THE Q6 CATHETER SEPARATED WITH THE WIRE PORTION IN THE SHEATH. THE PHYSICIAN USED A SNARE DEVICE TO SUCCESSFULLY REMOVE THE DETACHED CATHETER PORTION OF THE Q6. THE THROMBECTOMY PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 299614 | Q6 ASPIRATION CATHETER | Q6 ASPIRATION CATHETER | DQY | MIVI NEUROSCIENCE, INC. | Q6-69145-E | M18120007 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |