FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 8505822 · Received April 11, 2019

Report

Report Number
2024168-2019-02837
Event Type
Injury
Date Received
April 11, 2019
Date of Event
March 16, 2019
Report Date
May 8, 2019
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SI
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

INTERNAL FILE NUMBER - (B)(4). CORRECTION: LOT NUMBER UPDATED FROM 8110841 TO 8083141. EVALUATION SUMMARY: VISUAL AND FUNCTIONAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO DEPLOY THE NEEDLES WAS CONFIRMED AS A NEEDLE DEFLECTION/SUTURE RETRIEVAL ISSUE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE REPORTED TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4): DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE 2 ADDITIONAL PROGLIDE DEVICES REFERENCED ARE BEING FILED UNDER SEPARATE MEDWATCH REPORTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT SUTURE PLACEMENT WITH THREE PROGLIDE DEVICES VIA 14FR SHEATH WERE ATTEMPTED IN THE LEFT COMMON FEMORAL ARTERY USING THE PRECLOSE TECHNIQUE PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) PROCEDURE. REPORTEDLY, THE DEPLOYMENT OF THE NEEDLES PRIOR TO THE TAVI SHEATH INSERTION WAS UNSUCCESSFUL WITH THREE PROGLIDE DEVICES. TWO ADDITIONAL PROGLIDE DEVICES WERE USED FOR SUCCESSFUL SUTURE PLACEMENT USING THE PRECLOSE TECHNIQUE. THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED USING THE PRE-PLACED SUTURES FROM THE TWO ADDITIONAL PROGLIDE DEVICES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
296701 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 8083141

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention SHEATH: 14FRHEPARIN| SHEATH: 14FRHEPARIN