FDA Adverse Event Malfunction Summary report: N

FC 500 FLOW CYTOMETER

MDR report key: 8505283 · Received April 11, 2019

Report

Report Number
1061932-2019-01405
Event Type
Malfunction
Date Received
April 11, 2019
Date of Event
February 27, 2019
Report Date
April 11, 2019
Manufacturer
BECKMAN COULTER
Product Code
GKZ
UDI-DI
15099590572822
PMA / PMN Number
K071681
Removal / Correction Number
2050012-0108/2018-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) WAS ON SITE AND CONFIRMED WHAT THE CUSTOMER REPORTED. TO RESOLVE THE ISSUE THE FSE REPLACED THE TARPON AMP BOARD. BEC IS FILING AN MDR FOR THIS EVENT BASED ON THE FDA CLASSIFICATION OF THE 08 JAN 2018 URGENT MEDICAL DEVICE RECALL AS A CLASS I RECALL ON 20 NOV 2018 (RECALL NUMBER Z-0471-2019 FOR FC 500; RECALL NUMBER Z-0472-2019 FOR EPICS XL/XLMCL). BEC INTERNAL IDENTIFIER ¿ (B)(4).

Description of Event or Problem · 1

DURING PM ON CUSTOMER'S FC 500 FLOW CYTOMETER., THE FIELD SERVICE ENGINEER OBSERVED THE POT ON AMP BOARD AT FL1 IS NOT RESPONDING TO ADJUSTMENT. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298727 FC 500 FLOW CYTOMETER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER TN,FC500 MPL FLOW CYTOMETER 100-240V,50/60HZ 15099590572822