FDA Adverse Event Malfunction Summary report: N

AXIOM ICONOS R200

MDR report key: 8504197 · Received April 11, 2019

Report

Report Number
3004977335-2019-70845
Event Type
Malfunction
Date Received
April 11, 2019
Date of Event
February 18, 2019
Report Date
February 18, 2019
Manufacturer
SIEMENS HEALTHCARE GMBH
Product Code
JAA
PMA / PMN Number
K992660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE WAS INVESTIGATED BASED ON THE INFORMATION RECEIVED AND THE PROVIDED ERROR LOGS. THE REPLACED PARTS WERE NOT AVAILABLE FOR INVESTIGATION. THE ANALYSIS OF THE LOG FILES SHOWED NO INDICATION OF MALFUNCTION. THE ISSUE COULD HAVE BEEN CAUSED BY ONE OF THE TWO HARDWARE ISSUES - STUCK TILT BUTTON ON THE TABLE SIDE CONTROL OR A DE-CALIBRATED /DEFECTIVE JOYSTICK IN THE REMOTE CONTROL DESK. THE TWO COMPONENTS (TABLE SIDE CONTROL, MATERIAL NUMBER 7030773 AND CROSS SWITCH WITH POTI, MATERIAL NUMBER 7030765) HAVE BEEN ALREADY REPLACED AT THE CUSTOMER SITE. THE REPLACED PARTS WERE REQUESTED FOR FURTHER ANALYSIS; HOWEVER, THEY HAD BEEN DISPOSED LOCALLY. A DETAILED INVESTIGATION WAS THEREFORE NOT POSSIBLE. HOWEVER, BOTH AFFECTED PARTS SHOW LOW VALUES IN SPARE PART CONSUMPTION. THEREFORE, NO SYSTEMATIC ISSUE IS IDENTIFIED. THE PROBLEM DID NOT RECUR SINCE THE JOYSTICK AND TABLE SIDE CONTROL DESK WERE REPLACED.

Additional Manufacturer Narrative · 1

TABLE TILT CONTROLLER AND TABLE SIDE CONTROLLER HAVE BEEN ALREADY EXCHANGED AT THIS SITE. THE UNIT WORKS AS SPECIFIED. INVESTIGATION IS ON-GOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

SIEMENS BECAME AWARE OF UNINTENDED TABLE MOVEMENTS ON THE AXIOM ICONOS R200 UNIT. THE X-RAY TABLE TILTED FROM 90 DEGREE TO 0 DEGREE POSITION WITHOUT GIVEN COMMAND. SIEMENS IS NOT AWARE OF ANY PATIENT INVOLVEMENT IN THIS CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299094 AXIOM ICONOS R200 IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM JAA SIEMENS HEALTHCARE GMBH 5902775

Patients

Seq Age Sex Outcome Treatment
1 Other