FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 8497605 · Received April 9, 2019

Report

Report Number
9610877-2019-00235
Event Type
Malfunction
Date Received
April 9, 2019
Report Date
March 18, 2019
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOQ
PMA / PMN Number
K131028
Removal / Correction Number
NOT ISSUED YET
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). (EXEMPTION NUMBER E2015036).

Description of Event or Problem · 0

PENTAX OF AMERICA INITIATED FIELD CORRECTION 2017-008-C WHICH INCLUDED INSPECTION OF THE SUCTION ARM ON AFFECTED MODELS PURSUANT TO PREDEFINED INSPECTION CRITERIA ((B)(4)). THE OBJECTIVE OF THE INSPECTION WAS TO LOCATE PART C255-AB171 (SUCTION ARM) AND VERIFY IT IS NOT LOOSE. IF PART C255-AB171 (SUCTION ARM) IS FOUND LOOSE, THE DEVICE WAS CONSIDERED TO FAIL THE INSPECTION CRITERIA. THE CUSTOMER DEVICE WAS PREVIOUSLY RETURNED TO PENTAX MEDICAL FROM A CUSTOMER ON 14/MAR/2019. INSPECTION OF THE DEVICE WAS PERFORMED ON 18/MAR/2019 WHERE THE QUALITY CONTROL INSPECTOR FOUND THE FOLLOWING: SUCTION ARM LOOSE, INSERTION TUBE NON-PENTAX MATERIAL, LEAK AT PVE CONNECTOR, FAILED WET LEAK TEST, UMBILICAL CABLE NON-PENTAX MATERIAL, UP/ DOWN BRAKE LEVER CRACKED, DISTAL BODY CHIPPED, UP/ DOWN CONTROL KNOB/ LEVER CRACKED, LIGHTGUIDE PRONG SCRATCHED/ PITTED, FAILED DRY LEAK TEST, LIGHTGUIDE PRONG COVER GLASS SET SCRATCHED, LIGHT CARRYING BUNDLE BROKEN, LIGHTGUIDE PRONG COVER GLASS SET DENTED, CHEMICAL RESIDUE BUILDUP ON CONTROL BODY, PRIMARY OPERATION CHANNEL RESISTANCE, CCD CIRCUIT BOARD POOR SOLDER TECHNIQUE, HOLE IN # 2 REMOTE CONTROL BUTTON COVER, IMAGE BLACKOUT. INSPECTION OF THE SUCTION ARM WAS PERFORMED AND THE DEVICE FAILED THE INSPECTION CRITERIA. THE SCOPE'S REPAIRS TO BE PERFORMED WHERE THE LOOSE SUCTION ARM WILL BE CORRECTED, AND THE UNIT RETURNED TO THE USER UPON COMPLETION. PARTS REPLACED: O-RINGS AND SEALS, DISTAL END W/CCD-M PB-FREE/NTSC, INSERTION FLEX TUBE W/SEG PB-FREE, BENDING RUBBER, DISTAL ATTACHING PIN, SEGMENT STEEL BRAID, LIGHT GUIDE CABLE, ELECTRICAL CONNECTOR ASSY, LIGHT GUIDE PRONG IMP-1, LG PRONG INSULATION RING, LG PRONG ATTACHING NUT, LIGHT GUIDE PRONG WASHER IMP-1, LG COVER GLASS SET (LENS TYPE), FRICTION LEVER ASSY, ANGLE LEVER ASSY, INSERTION/S-NIPPLE ATTACHING SCREW, SUCTION CONNECTION TUBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288465 PENTAX VIDEO BRONCHOSCOPE EOQ HOYA CORPORATION PENTAX TOKYO OFFICE EB-1170K

Patients

Seq Age Sex Outcome Treatment
1