PENTAX
Report
- Report Number
- 9610877-2019-00235
- Event Type
- Malfunction
- Date Received
- April 9, 2019
- Report Date
- March 18, 2019
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOQ
- PMA / PMN Number
- K131028
- Removal / Correction Number
- NOT ISSUED YET
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). (EXEMPTION NUMBER E2015036).
PENTAX OF AMERICA INITIATED FIELD CORRECTION 2017-008-C WHICH INCLUDED INSPECTION OF THE SUCTION ARM ON AFFECTED MODELS PURSUANT TO PREDEFINED INSPECTION CRITERIA ((B)(4)). THE OBJECTIVE OF THE INSPECTION WAS TO LOCATE PART C255-AB171 (SUCTION ARM) AND VERIFY IT IS NOT LOOSE. IF PART C255-AB171 (SUCTION ARM) IS FOUND LOOSE, THE DEVICE WAS CONSIDERED TO FAIL THE INSPECTION CRITERIA. THE CUSTOMER DEVICE WAS PREVIOUSLY RETURNED TO PENTAX MEDICAL FROM A CUSTOMER ON 14/MAR/2019. INSPECTION OF THE DEVICE WAS PERFORMED ON 18/MAR/2019 WHERE THE QUALITY CONTROL INSPECTOR FOUND THE FOLLOWING: SUCTION ARM LOOSE, INSERTION TUBE NON-PENTAX MATERIAL, LEAK AT PVE CONNECTOR, FAILED WET LEAK TEST, UMBILICAL CABLE NON-PENTAX MATERIAL, UP/ DOWN BRAKE LEVER CRACKED, DISTAL BODY CHIPPED, UP/ DOWN CONTROL KNOB/ LEVER CRACKED, LIGHTGUIDE PRONG SCRATCHED/ PITTED, FAILED DRY LEAK TEST, LIGHTGUIDE PRONG COVER GLASS SET SCRATCHED, LIGHT CARRYING BUNDLE BROKEN, LIGHTGUIDE PRONG COVER GLASS SET DENTED, CHEMICAL RESIDUE BUILDUP ON CONTROL BODY, PRIMARY OPERATION CHANNEL RESISTANCE, CCD CIRCUIT BOARD POOR SOLDER TECHNIQUE, HOLE IN # 2 REMOTE CONTROL BUTTON COVER, IMAGE BLACKOUT. INSPECTION OF THE SUCTION ARM WAS PERFORMED AND THE DEVICE FAILED THE INSPECTION CRITERIA. THE SCOPE'S REPAIRS TO BE PERFORMED WHERE THE LOOSE SUCTION ARM WILL BE CORRECTED, AND THE UNIT RETURNED TO THE USER UPON COMPLETION. PARTS REPLACED: O-RINGS AND SEALS, DISTAL END W/CCD-M PB-FREE/NTSC, INSERTION FLEX TUBE W/SEG PB-FREE, BENDING RUBBER, DISTAL ATTACHING PIN, SEGMENT STEEL BRAID, LIGHT GUIDE CABLE, ELECTRICAL CONNECTOR ASSY, LIGHT GUIDE PRONG IMP-1, LG PRONG INSULATION RING, LG PRONG ATTACHING NUT, LIGHT GUIDE PRONG WASHER IMP-1, LG COVER GLASS SET (LENS TYPE), FRICTION LEVER ASSY, ANGLE LEVER ASSY, INSERTION/S-NIPPLE ATTACHING SCREW, SUCTION CONNECTION TUBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 288465 | PENTAX | VIDEO BRONCHOSCOPE | EOQ | HOYA CORPORATION PENTAX TOKYO OFFICE | EB-1170K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |