IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM
Report
- Report Number
- 0002023141-2019-00148
- Event Type
- Malfunction
- Date Received
- April 9, 2019
- Date of Event
- March 11, 2019
- Report Date
- July 10, 2019
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- PMA / PMN Number
- K061410
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ONE IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM (TSVB13) WAS RETURNED FOR INVESTIGATION ATTACHED TO A MATING MOUNT. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED WEAR AND DEBRIS ABOUT THE IMPLANT THREADS AND VENT. FUNCTIONAL TESTING WAS PERFORMED FOR THE RETURNED PRODUCT AND IT WAS DETERMINED THAT THE SCREW COULD BE EASILY REMOVED, HOWEVER THE MOUNT HEX TOOK EXCESSIVE FORCE TO REMOVE. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 63725066. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (63725066) FOR SIMILAR CAUSE AND NO OTHER COMPLAINT WAS IDENTIFIED. APRIL POST MARKET TRENDING WAS REVIEWED AND THERE WERE NO NEGATIVE TRENDS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT(STUCK MOUNT) OR DEVICE (TSVB13). THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED. A DEFINITIVE CAUSE COULD NOT BE IDENTIFIED; HOWEVER, THE PROBABLE CAUSE MAY BE ASSOCIATED TO EXCESSIVE LOADING ON MOUNT/IMPLANT ASSEMBLY. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D4: UNIQUE IDENTIFIER (UDI) NUMBER: (B)(4).
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT 2031527-2019-00177 PMA/510(K) NUMBER: K011028, K013227.
IT WAS REPORTED THAT A MOUNT COULD NOT DISENGAGE FROM IMPLANT (TSVB13). IMPLANT WAS REMOVED AND ANOTHER IMPLANT WAS PLACED. TOOTH LOCATION 29.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 289293 | IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 63725066 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |