FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM

MDR report key: 8496700 · Received April 9, 2019

Report

Report Number
0002023141-2019-00148
Event Type
Malfunction
Date Received
April 9, 2019
Date of Event
March 11, 2019
Report Date
July 10, 2019
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K061410
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ONE IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM (TSVB13) WAS RETURNED FOR INVESTIGATION ATTACHED TO A MATING MOUNT. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED WEAR AND DEBRIS ABOUT THE IMPLANT THREADS AND VENT. FUNCTIONAL TESTING WAS PERFORMED FOR THE RETURNED PRODUCT AND IT WAS DETERMINED THAT THE SCREW COULD BE EASILY REMOVED, HOWEVER THE MOUNT HEX TOOK EXCESSIVE FORCE TO REMOVE. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 63725066. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (63725066) FOR SIMILAR CAUSE AND NO OTHER COMPLAINT WAS IDENTIFIED. APRIL POST MARKET TRENDING WAS REVIEWED AND THERE WERE NO NEGATIVE TRENDS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT(STUCK MOUNT) OR DEVICE (TSVB13). THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED. A DEFINITIVE CAUSE COULD NOT BE IDENTIFIED; HOWEVER, THE PROBABLE CAUSE MAY BE ASSOCIATED TO EXCESSIVE LOADING ON MOUNT/IMPLANT ASSEMBLY. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D4: UNIQUE IDENTIFIER (UDI) NUMBER: (B)(4).

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT 2031527-2019-00177 PMA/510(K) NUMBER: K011028, K013227.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MOUNT COULD NOT DISENGAGE FROM IMPLANT (TSVB13). IMPLANT WAS REMOVED AND ANOTHER IMPLANT WAS PLACED. TOOTH LOCATION 29.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289293 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM DENTAL IMPLANT DZE ZIMMER DENTAL 63725066

Patients

Seq Age Sex Outcome Treatment
1