FDA Adverse Event Malfunction Summary report: N

PROSOUND F37

MDR report key: 8496094 · Received April 9, 2019

Report

Report Number
8030405-2019-00002
Event Type
Malfunction
Date Received
April 9, 2019
Date of Event
March 15, 2019
Report Date
April 9, 2019
Manufacturer
HITACHI MEDICAL CORPORATION
Product Code
IYN
PMA / PMN Number
K122341
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE PROSOUND F37 IS A DIAGNOSTIC ULTRASOUND SYSTEM. HITACHI RECEIVED A COMPLAINT ON (B)(6) 2019 REGARDING THE F37 ULTRASOUND SYSTEM. THE SITE REPORTED A DISCREPANCY BETWEEN THE EFW (ESTIMATED FETAL WEIGHT) AND PERCENTILE NUMBER TO THE NUMBER THAT IS GRAPHED IN THE OB REPORT. HITACHI APPLICATIONS CONTACTED THE CUSTOMER AND REVIEWED THEIR PRESET SETTINGS. PRESET SETTINGS INCLUDE SEVERAL DIFFERENT CALCULATIONS (I.E. HADLOCK, DOUBILET), WHICH THE CUSTOMERS CAN SELECT. THE SETTINGS DICTATE THE CALCULATION TABLE THAT THE SYSTEM WILL USE TO PERFORM THE CALCULATION AND ALSO AFFECT THE LABELS FOR THE GRAPH AND WEIGHT ESTIMATES OF THE OB REPORTS. THE SETTINGS CAN BE MODIFIED BY THE CUSTOMER AT WILL. HITACHI APPLICATIONS ASSISTED THE CUSTOMER IN CHANGING THEIR PRESET SETTINGS TO HADLOCK5 FOR THE APPROPRIATE CALCULATION AND GROWTH TABLE. THE SYSTEM IS OPERATING WITHIN SPECIFICATIONS AFTER APPLYING THE CHANGE. THE ROOT CAUSE OF THE PROBLEM IS USER ERROR. NOTE: THE PROSOUND F37 WAS MANUFACTURED IN (B)(4) BUT MANUFACTURING HAS BEEN MOVED TO ANOTHER SITE.

Description of Event or Problem · 1

ON (B)(6) 2019, HITACHI RECEIVED A COMPLAINT REGARDING THE F37 ULTRASOUND SYSTEM. THE SITE REPORTED A DISCREPANCY BETWEEN THE EFW (ESTIMATED FETAL WEIGHT) AND PERCENTILE NUMBER TO THE NUMBER THAT IS GRAPHED IN THE OB REPORT. THE SITE DID NOT REPORT ANY OTHER ISSUE WITH THE ULTRASOUND SYSTEM OR OB REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
287514 PROSOUND F37 PROSOUND F37 IYN HITACHI MEDICAL CORPORATION PROSOUND F37 N/A

Patients

Seq Age Sex Outcome Treatment
1 Other