BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE
Report
- Report Number
- 1024879-2019-00709
- Event Type
- Malfunction
- Date Received
- April 8, 2019
- Date of Event
- March 18, 2019
- Report Date
- April 23, 2019
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- FMI
- UDI-DI
- 50382903686082
- PMA / PMN Number
- K982541
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: AS NO SAMPLES OR PHOTOS WERE RECEIVED FOR EVALUATION, THE CUSTOMER'S INDICATED FAILURE MODE WAS NOT OBSERVED BY BD. ROOT CAUSE DESCRIPTION: AS THERE WAS NO SAMPLE OR PHOTO AVAILABLE FOR EVALUATION, A ROOT CAUSE COULD NOT BE DETERMINED. RATIONALE: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT A BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLES HAD SAFETIES BREAKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "MATERIAL NO. 368608, BATCH NO. 9010767. IT WAS REPORTED THAT THE SAFETY IS BREAKING ON 10 NEEDLES IN HER BOX.. CUSTOMER IS REPORTING THE SAFETY BREAKING ON 10 NEEDLES IN HER BOX.. ADDITIONAL INFO AWARE COMMENTS; WHAT IS THE DATE OF OCCURRENCE? (B)(6) 2019. DID ALL 10 HAPPEN ON THE SAME DAY? IF NOT BECAUSE SPECIFIC THE NUMBER OF OCCURRENCES FOR EACH DATE. 1 HAPPENED THIS DAY. THROUGHOUT USING THESE NEEDLES, WE HAVE HAD OTHERS THAT I FORWARDED TO MEDLINE. WHEN DID THE SAFETY BREAK? (BEFORE USE, DURING USE, AFTER USE) IT BROKE WHEN CLOSING THE SAFETY (SAME THING WITH THE PREVIOUS OCCURRENCES). COULD YOU EXPLAIN INTO FURTHER DETAIL, HOW THE SAFETY IS BREAKING? IT BREAKS WHEN THE SAFETY IS GETTING CLOSED. DO YOU HAVE ANY PATIENT IDENTIFIERS FOR THESE 10 OCCURRENCES? HAS HAPPENED WHEN STUDENTS ARE JUST PRACTICING CLOSING SAFETY'S AND AFTER A BLOOD DRAW. DO YOU HAVE ANY UNUSED PRODUCT (5-10 SAMPLES) THAT CAN BE SENT IN FOR INVESTIGATION? YES, WE DO. CAN YOU PLEASE SEND DETAILS FOR MAILING BACK TO YOU? IS THERE A WAY WE CAN GET A REPLACEMENT FOR THE AFFECTED LOT? WERE THERE ANY NEEDLE STICKS? IF SO, WHAT WAS THE MEDICAL INTERVENTION? NO."
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT A BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLES HAD SAFETIES BREAKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "MATERIAL NO. 368608 BATCH NO. 9010767''. IT WAS REPORTED THAT THE SAFETY IS BREAKING ON 10 NEEDLES IN HER BOX.. CUSTOMER IS REPORTING THE SAFETY BREAKING ON 10 NEEDLESS IN HER BOX.. ADDITIONAL INFO AWARE COMMENTS, WHAT IS THE DATE OF OCCURRENCE? (B)(6) 2019. DID ALL 10 HAPPEN ON THE SAME DAY? IF NOT BECAUSE SPECIFIC THE NUMBER OF OCCURRENCES FOR EACH DATE. 1 HAPPENED THIS DAY. THROUGHOUT USING THESE NEEDLES, WE HAVE HAD OTHERS THAT I FORWARDED TO MEDLINE. WHEN DID THE SAFETY BREAK? (BEFORE USE, DURING USE, AFTER USE) IT BROKE WHEN CLOSING THE SAFETY (SAME THING WITH THE PREVIOUS OCCURRENCES). COULD YOU EXPLAIN INTO FURTHER DETAIL, HOW THE SAFETY IS BREAKING? IT BREAKS WHEN THE SAFETY IS GETTING CLOSED. DO YOU HAVE ANY PATIENT IDENTIFIERS FOR THESE 10 OCCURRENCES? HAS HAPPENED WHEN STUDENTS ARE JUST PRACTICING CLOSING SAFETY AND AFTER A BLOOD DRAW. DO YOU HAVE ANY UNUSED PRODUCT (5-10 SAMPLES) THAT CAN BE SENT IN FOR INVESTIGATION? YES, WE DO. CAN YOU PLEASE SEND DETAILS FOR MAILING BACK TO YOU? IS THERE A WAY WE CAN GET A REPLACEMENT FOR THE AFFECTED LOT? WERE THERE ANY NEEDLE STICKS? IF SO, WHAT WAS THE MEDICAL INTERVENTION? NO."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 283434 | BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON, DICKINSON & CO., (BD) | 9010767 | 50382903686082 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |