FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 8491911
·
Received April 8, 2019
Report
- Report Number
- 9611109-2019-00229
- Event Type
- Malfunction
- Date Received
- April 8, 2019
- Date of Event
- March 11, 2019
- Report Date
- April 8, 2019
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE AND COULD CONFIRM THE REPORTED ISSUE. HE TRACED THE FAILURE TO THE DAMAGED COMPRESSOR OF THE COOLING SYSTEM. THE DEVICE WAS TAKEN OUT OF SERVICE. CORRECTIVE ACTIONS ARE IN PLACE FOR THIS ISSUE.
Description of Event or Problem · 1
LIVANOVA (B)(4) RECEIVED A REPORT THAT DURING PRIMING THE HEATER-COOLER 3T TRIGGERED THE FUSES WHEN TURNED ON. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 286108 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |