FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 8491911 · Received April 8, 2019

Report

Report Number
9611109-2019-00229
Event Type
Malfunction
Date Received
April 8, 2019
Date of Event
March 11, 2019
Report Date
April 8, 2019
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE AND COULD CONFIRM THE REPORTED ISSUE. HE TRACED THE FAILURE TO THE DAMAGED COMPRESSOR OF THE COOLING SYSTEM. THE DEVICE WAS TAKEN OUT OF SERVICE. CORRECTIVE ACTIONS ARE IN PLACE FOR THIS ISSUE.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT DURING PRIMING THE HEATER-COOLER 3T TRIGGERED THE FUSES WHEN TURNED ON. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
286108 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1