FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX

MDR report key: 8488176 · Received April 5, 2019

Report

Report Number
1920898-2019-00322
Event Type
Malfunction
Date Received
April 5, 2019
Date of Event
March 20, 2019
Report Date
June 4, 2019
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: CUSTOMER RETURNED PHOTOS OF A 1/2CC, 12.7MM, 29G SYRINGE WITH THE BLISTER PACK FROM LOT # 7240991. CUSTOMER STATES THAT THE BARREL IS DESTROYED. THE ATTACHED PHOTOS WERE EXAMINED AND EXHIBITED CRACKS IN THE BARREL RANGING FROM THE 0-15 UNIT MARKINGS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7240991. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE THREE (3) NOTIFICATIONS [200720172, 200720171, 200706852] NOTED FOR DAMAGED BARREL. THERE WERE TWO (2) NOTIFICATIONS [200705296, 200704761] NOTED FOR DAMAGED TIPS. THERE WERE TEN (10) NOTIFICATIONS [200719948, 200720205, 200720038, 200720172, 200719960, 200719959, 200720006, 200704761, 200706359, 200706592] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. POSSIBLE ROOT CAUSES: DEFECT COULD HAVE OCCURRED AT THE PREP DIAL OF THE METRO ASSEMBLY MACHINE LOSS OF BACK PRESSURE ON THE IN-FEED RAIL MAY CAUSE THE BARREL TO BECOME PINCHED AT THE RAIL / PREP DIAL JUNCTION. THE TRIP GATE EYE SENSOR SHOULD DETECT LOW RAIL CONDITIONS AND ACTIVATE THE TRIP GATE TO PREVENT THE BARRELS FROM BEING LOADED INTO THE PREP DIAL AND POSSIBLY JAMMING AT THAT LOCATION; INFEED DIAL AT BARREL PRINTERS; LOSS OF BACK PRESSURE AT INFEED RAIL ALSO AT FORM FILL & SEAL.

Description of Event or Problem · 0

IT WAS REPORTED THE SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX BARREL WAS FOUND DAMAGED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: "THE SYRINGE BARREL ARE DESTROYED".

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THE SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX BARREL WAS FOUND DAMAGED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "THE SYRINGE BARREL ARE DESTROYED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282082 SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 7240991

Patients

Seq Age Sex Outcome Treatment
1 Other