FDA Adverse Event Malfunction Summary report: N

O2

MDR report key: 8485018 · Received April 4, 2019

Report

Report Number
3004785967-2019-00576
Event Type
Malfunction
Date Received
April 4, 2019
Date of Event
March 29, 2019
Report Date
May 1, 2019
Manufacturer
MEDTRONIC NAVIGATION, INC (LITTLETON)
Product Code
OWB
UDI-DI
00643169639683
PMA / PMN Number
K151000
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. TESTING REVEALED THAT ETHERNET PATCH CORD TO STEALTH THAT PREVENTED CONNECTION, ALSO FOUND THAT IOR WAS TURNED ENABLED ON O ARM WHICH MAY CAUSE INTERMITTENT ISSUES. UTILIZED WORKING PATCH CORD, DISABLED IOR IN SYSTEM FEATURES. VERIFIED OPERATION WITH S7 AND TOLD TECHS OF OPERATION WITHOUT IOR DROP DOWN MENU. ACCOUNT DOES NOT HAVE IOR SYSTEM INSTALLED. THE IMAGING SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: PATIENT INFORMATION UPDATED. PATIENT WEIGHT WAS UNAVAILABLE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION VIA AN FIELD SERVICE ENGINEER (FSE) FROM A RADIATION TECHNICIAN (RT) REGARDING AN IMAGING SYSTEM BEING USED FOR A PROCEDURE. THE RT INDICATED THAT THE IMAGING SYSTEM HAD A CONNECTION ISSUE WITH THE NAVIGATION SYSTEM IN THAT THE SYSTEMS WERE NOT CONNECTING. THE NETWORK CABLE WAS DISCONNECTED FROM THE NAVIGATION SYSTEM AND RECONNECTED WHICH RESOLVED THE ISSUE. THERE WAS NO IMPACT TO PATIENT OUTCOME.

Description of Event or Problem · 1

THE ISSUE WAS DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. THE ISSUE EXTENDED THE SURGICAL TIME BY LESS THAN 1 HOUR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277160 O2 INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB MEDTRONIC NAVIGATION, INC (LITTLETON) BI70002000 00643169639683

Patients

Seq Age Sex Outcome Treatment
1 76 YR