FDA Adverse Event Injury Summary report: N

M2A-MAGNUM PF CUP 52ODX46ID

MDR report key: 8483969 · Received April 4, 2019

Report

Report Number
0001825034-2019-01568
Event Type
Injury
Date Received
April 4, 2019
Date of Event
January 10, 2019
Report Date
March 29, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWA
PMA / PMN Number
K042037
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT SAMPLE WAS RETURNED AND EVALUATED AGAINST THE REPORTED EVENT. VISUAL INSPECTION FOUND THE RETURNED CUP FOUND THAT THE INNER RADIUS OF THE CUP IS SCRATCHED. THE INNER RADIUS IS COVERED IN A DRIED YELLOW LIQUID. YELLOW AND BLACK DEBRIS REMAIN STUCK TO THE POROUS COATING. ADDITIONAL INFORMATION DOES NOT CHANGE THE ROOT CAUSE OF THE PREVIOUS INVESTIGATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.  REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MW5084465.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN AND ELEVATED METAL ION LEVELS APPROXIMATELY 11 YEARS POST IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: CATALOG NUMBER: 157446 LOT NUMBER:263600 BRAND NAME: M2A MAGNUM HEAD, CATALOG NUMBER: 139259 LOT NUMBER:070130 BRAND NAME: M2S MAGNUM TAPER INSERT, CATALOG NUMBER: X180311 LOT NUMBER:752740 BRAND NAME: BI-METRIC STEM. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2019-01569. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. DEVICE REMAINS IMPLANTED IN THE PATIENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT IS EXPERIENCING RIGHT HIP PAIN AND ELEVATED METAL ION LEVELS APPROXIMATELY 11 YEARS POST IMPLANTATION. HOWEVER, NO REVISION HAS BEEN REPORTED TO DATE. ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276562 M2A-MAGNUM PF CUP 52ODX46ID PROSTHESIS, HIP KWA ZIMMER BIOMET, INC. N/A 782650 

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| O| R