FDA Adverse Event Malfunction Summary report: N

VOLISTA

MDR report key: 8483022 · Received April 4, 2019

Report

Report Number
9710055-2019-00109
Event Type
Malfunction
Date Received
April 4, 2019
Date of Event
March 29, 2019
Report Date
July 12, 2019
Manufacturer
MAQUET SAS
Product Code
FSY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET SAS, PARC DE LIMÈRE, AVENUE DE LA POMME DE PI ORLÉANS CEDEX 2, FRANCE 45074 EXEMPTION # E2018005. GETINGE USA SALES, LLC, 45 BARBOUR POND DRIVE, WAYNE, NJ 07470. CONTACT PERSON: (B)(4). THE ISSUE IS STILL BEING INVESTIGATED BY MANUFACTURING SITE.

Additional Manufacturer Narrative · 0

MAQUET SAS BECAME AWARE OF A CUSTOMER¿S PROBLEM WITH THE VOLISTA DEVICE. AS IT WAS STATED BY THE TECHNICIAN, THERE WERE CRACKS IN THE PAINT OBSERVED, SPECIFICALLY AT THE ARC OF THE FORK. THERE WAS NO INJURY TO A PATIENT REPORTED, HOWEVER IT WAS DECIDED TO REPORT THIS ISSUE BASED ON THE POTENTIAL AND IN ABUNDANCE OF CAUTION, AS ANY PARTICLE FALLING FROM THE DEVICE INTO THE STERILE FIELD MIGHT BE A SOURCE OF CONTAMINATION. WHEN REVIEWING SIMILAR REPORTABLE EVENTS REGISTERED FOR VOLISTA SURGICAL LIGHT WE WERE ABLE TO FIND SEVERAL SIMILAR ISSUES COMPARED TO THE PROBLEM INVESTIGATED HEREIN. IN NONE OF THE COMPLAINTS A SERIOUS INJURY OR DEATH OCCURRED. BASED ON THE INFORMATION COLLECTED TO DATE IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE SURGICAL LIGHT DID NOT MEET THE MANUFACTURER¿S SPECIFICATION. THE MOST PROBABLE ROOT CAUSE THAT THE CRACKS APPEAR WAS THE DETACHMENT OF THE BONDED PARTS OF FORK. IT WAS ESTABLISHED THAT THE CRACKS WERE LOCATED ONLY ON THE OUTER COATING AND THERE WAS NO IMPACT TO THE INTERNAL STRUCTURE OF THE FORK ASSEMBLY. THE BREAKS ON THE COATING CORRESPOND TO THE EXCESSIVE GAP BETWEEN TWO MECHANICAL PARTS UNDER THE OUTER MATERIAL. WE HAVE NO INFORMATION REGARDING THE EXACT TIME WHEN THE DEFECT FIRST APPEARED OR IF IT WAS BEING USED FOR PATIENT TREATMENT IN THE TIME WHEN THE EVENT OCCURRED. A CORRECTIVE/PREVENTIVE ACTION INVESTIGATION INTO THE ISSUE HAS REVEALED THAT THE ISSUE MAY BE THE RESULT OF SEVERAL FACTORS, SUMMARIZED AS ISSUES WITH THE PREVIOUS UNDERSTANDING OF POST-GLUING ASPECTS. A DESIGN CHANGE IMPROVED THE ASSEMBLING METHODS FROM GLUE-BONDING TO WELDING OF THE PARTS IN PRODUCTION, SINCE BEGINNING OF 2017. IN ADDITION, THE DEVICE WAS INCLUDED IN THE SCOPE OF FIELD ACTION MSA-2019-001-IU AND IT WAS DECIDED THAT THE DEFECTIVE PART WILL BE REPLACED, THEREFORE GETINGE DOES NOT PROPOSE ANY OTHER ACTION AT THIS TIME.

Description of Event or Problem · 0

ON 29TH MARCH, 2019 MAQUET SAS BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS- VOLISTA. AS IT WAS STATED, THE DEVICE HAS A CRACK ON THE FORK WHICH MAY LEAD TO PAINT PARTICLES FALLING OFF. THERE WAS NO INJURY REPORTED HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PAINT PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MIGHT BE A SOURCE OF CONTAMINATION. MANUFACTURER REFERENCE NUMBER (B)(4).

Description of Event or Problem · 0

MANUFACTURER REFERENCE NUMBER 2019-63115.

Additional Manufacturer Narrative · 1

THE ISSUE IS STILL BEING INVESTIGATED BY MANUFACTURING SITE. (B)(4). EXEMPTION # E2018005. (B)(4).

Additional Manufacturer Narrative · 1

THE ISSUE IS STILL BEING INVESTIGATED BY MANUFACTURING SITE. (B)(4). EXEMPTION # E2018005. (B)(4).

Additional Manufacturer Narrative · 1

THE ISSUE IS STILL BEING INVESTIGATED BY MANUFACTURING SITE. (B)(4). EXEMPTION # E2018005. (B)(4).

Description of Event or Problem · 1

MANUFACTURER REFERENCE NUMBER (B)(4).

Description of Event or Problem · 1

MANUFACTURER REFERENCE NUMBER (B)(4).

Description of Event or Problem · 1

MANUFACTURER REFERENCE NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277544 VOLISTA LIGHT, SURGICAL, CEILING MOUNTED FSY MAQUET SAS

Patients

Seq Age Sex Outcome Treatment
1