STEM: AMISTEM H HA COATED STD STEM SIZE 3
Report
- Report Number
- 3005180920-2019-00227
- Event Type
- Injury
- Date Received
- April 3, 2019
- Date of Event
- March 1, 2019
- Report Date
- April 3, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030804090
- PMA / PMN Number
- K093944
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 01 APRIL 2019: LOT 170012: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22-JUN-2017. EXPIRATION DATE: 2022-06-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VISUAL INSPECTION PERFORMED BY MEDACTA R&D PROJECT MANAGER: VISUAL INSPECTION SHOWS HA LAYER NOT ABSORBED ON THE PROXIMAL REGION. THIS IS IN LINE WITH THE LOOSENING DIAGNOSIS. CLINICAL EVALUATION PERFORMED BY MEDACTA MEDICAL AFFAIRS MANAGER: FEMORAL COMPONENT REVISION SURGERY OCCURRED 2 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) MAN. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES IN GRUEN ZONE 1 AND SIGNS OF STRESS SHIELDING. NO INFORMATION CONCERNING COMORBIDITIES AND GENERAL PATIENT HEALTH IS AVAILABLE. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS THA AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS EVENT CANNOT BE DETERMINED.
REVISION SURGERY PERFORMED 1 YEAR AND 11 MONTHS AFTER THE PRIMARY DUE TO STEM LOOSENING. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272541 | STEM: AMISTEM H HA COATED STD STEM SIZE 3 | UNCEMENTED HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 170012 | 07630030804090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention |