FDA Adverse Event
Malfunction
Summary report: N
HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT
MDR report key: 8478384
·
Received April 3, 2019
Report
- Report Number
- 3005180920-2019-00233
- Event Type
- Malfunction
- Date Received
- April 3, 2019
- Date of Event
- March 5, 2019
- Report Date
- April 3, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- UDI-DI
- 07630040742115
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
BATCH REVIEW PERFORMED ON 01 APRIL 2019: LOT 1854515: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-DEC-2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ANOTHER SIMILAR EVENT HAS BEEN REPORTED ON THIS SAME LOT (COMPLAINT (B)(4) [MDR 2019-0017]). VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER: THE SPRING BALL PLUNGER HAS COME APART. IT IS POSSIBLE THAT THIS OCCURRENCE COULD HAVE BEEN INFLUENCED BY VARIATION IN PRODUCTION/ASSEMBLY HOWEVER THIS CANNOT BE CONFIRMED.
Description of Event or Problem · 1
DURING SURGERY, THE LOCKING BALL FELL OUT FROM THE HANDLE. THE SURGEON USED A STRAIGHT BROACH HANDLE TO COMPLETE THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273458 | HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT | SURGICAL INSTRUMENT FOR HIP SURGERY | LXH | MEDACTA INTERNATIONAL SA | 1854515 | 07630040742115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |