FDA Adverse Event Malfunction Summary report: N

HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT

MDR report key: 8478384 · Received April 3, 2019

Report

Report Number
3005180920-2019-00233
Event Type
Malfunction
Date Received
April 3, 2019
Date of Event
March 5, 2019
Report Date
April 3, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
UDI-DI
07630040742115
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 01 APRIL 2019: LOT 1854515: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-DEC-2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ANOTHER SIMILAR EVENT HAS BEEN REPORTED ON THIS SAME LOT (COMPLAINT (B)(4) [MDR 2019-0017]). VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER: THE SPRING BALL PLUNGER HAS COME APART. IT IS POSSIBLE THAT THIS OCCURRENCE COULD HAVE BEEN INFLUENCED BY VARIATION IN PRODUCTION/ASSEMBLY HOWEVER THIS CANNOT BE CONFIRMED.

Description of Event or Problem · 1

DURING SURGERY, THE LOCKING BALL FELL OUT FROM THE HANDLE. THE SURGEON USED A STRAIGHT BROACH HANDLE TO COMPLETE THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273458 HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT SURGICAL INSTRUMENT FOR HIP SURGERY LXH MEDACTA INTERNATIONAL SA 1854515 07630040742115

Patients

Seq Age Sex Outcome Treatment
1 Other