FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 8478317 · Received April 3, 2019

Report

Report Number
9617229-2019-02261
Event Type
Injury
Date Received
April 3, 2019
Date of Event
February 12, 2019
Report Date
June 4, 2019
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: CREASES FLAT AND OPENING CURVED ON PATCH. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: OBSERVED VOID >1 MM IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA AND SHARP OPENING ON PATCH. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP OPENING ON PATCH (SHELL BOND EDGE) DUE TO AN UNIDENTIFIED (TEAR) OPENING. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 1

DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271148 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 2415263

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention