SYSMEX CS-2100I
Report
- Report Number
- 1000515253-2019-00009
- Event Type
- Malfunction
- Date Received
- April 3, 2019
- Date of Event
- February 27, 2019
- Report Date
- August 22, 2019
- Manufacturer
- SYSMEX CORPORATION, I SQUARE
- Product Code
- JPA
- UDI-DI
- 04987562409471
- PMA / PMN Number
- K162688
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THE SYSMEX CS-2000I/2100I INSTRUCTIONS FOR USE (IFU), CHAPTER 2 - SAFETY INFORMATION, SECTION 2.2 - GENERAL INFORMATION WARNS THE USER: SHOULD THE INSTRUMENT EMIT ANY UNUSUAL ODORS OR SMOKE, TURN THE MAIN SWITCH OFF IMMEDIATELY AND UNPLUG THE POWER CABLE. CONTACT YOUR LOCAL TECHNICAL REPRESENTATIVE. THE USER APPROPRIATELY TURNED OFF THE ANALYZER, DISCONNECTED THE POWER, AND CONTACTED SIEMENS HEALTHCARE DIAGNOSTICS. THE SE REPLACED THE MOTOR BLOCK NO. 126 ASSY. WITHOUT RESOLUTION. THE SE RETURNED TO THE CUSTOMER SITE AND REPLACED PCB NO. 1279, WHICH RESOLVED THE ISSUE. THE ANALYZER COVER AND MOTOR ENCODER DISC ARE MADE OF FLAME RESISTANT MATERIALS, BUT EXCESSIVE HEAT FROM A MALFUNCTION POSES A RISK OF INHALATION OF SMOKE AND/OR HARMFUL VAPORS FROM SCORCHED MATERIALS. THE AFFECTED PARTS ARE COVERED WITH A METAL SHEET AND NOT ACCESSIBLE TO THE USER, REDUCING POTENTIAL HARM. SYSMEX CORPORATION JAPAN (S-CORP) REQUESTED THE PARTS FOR INVESTIGATION.
SYSMEX CORPORATION (B)(4) (S-CORP) COMPLETED THE INVESTIGATION. S-CORP DETERMINED A SHORT CIRCUIT FROM EITHER THE PRINTED CIRCUIT BOARD (PCB) NO. 1279 OR THE STEPPING MOTOR (A COMPONENT OF MOTOR BLOCK ASSEMBLY NO. 126) CAUSED THE EVENT. THESE PARTS ARE CONNECTED; THEREFORE, S-CORP WAS UNABLE TO DETERMINE THE ORIGIN OF THE SHORT CIRCUIT. S-CORP DETERMINED THIS IS AN ISOLATED EVENT. THERE WAS NO SYSTEMIC DEFICIENCY IDENTIFIED, AS NO SIMILAR ISSUE WAS REPORTED. NO FURTHER CORRECTIVE ACTION WILL BE TAKEN.
A USER IN (B)(6) REPORTED A BURNT ODOR EMANATING FROM THE ANALYZER. THE USER IMMEDIATELY TURNED OFF THE ANALYZER, DISCONNECTED THE POWER, AND CONTACTED SIEMENS HEALTHCARE DIAGNOSTICS. A SIEMENS SERVICE ENGINEER (SE) WAS DISPATCHED THE SAME DAY AND FOUND THE CP UNIT DRIVE OF MOTOR BLOCK ASSEMBLY NO. 126 WAS BURNT, PARTS OF THE MOTOR ENCODER DISC WERE MELTED, A FUSE ON THE POWER SUPPLY (WIRING CORD NO. 5391) WAS BLOWN, AND PRINTED CIRCUIT BOARD (PCB) NO. 1279 WAS DAMAGED. SMOKE WAS OBSERVED; HOWEVER, NO OPEN FIRE WAS OBSERVED. NO USERS SOUGHT MEDICAL ATTENTION DUE TO THE BURNING ODOR. NO USER OR PATIENT HARM OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272324 | SYSMEX CS-2100I | AUTOMATED COAGULATION ANALYZER | JPA | SYSMEX CORPORATION, I SQUARE | CS-2100I | 04987562409471 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |