FDA Adverse Event Malfunction Summary report: Y

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 8477894 · Received April 3, 2019

Report

Report Number
2531779-2019-02387
Event Type
Malfunction
Date Received
April 3, 2019
Report Date
March 31, 2019
Manufacturer
ANIMAS LLC
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION:IN QUARTER 2 OF 2019, THERE WERE 2 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.

Additional Manufacturer Narrative · 1

OF THE 20 ALLEGATIONS OF A MALFUNCTION, 9 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 3 WERE FOUND TO BE MALFUNCTIONING DUE TO CRACKED OR TORN TRANSCEIVER FLEX. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.

Description of Event or Problem · 1

THIS REPORT SUMMARIZES <NOE> 20</NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 4-70 YEARS OF AGE AND BETWEEN 100 AND 247 POUNDS. OF THE REPORTED PATIENTS, 7 WERE MALE, 13 WERE FEMALE, AND 0 WERE UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273077 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS LLC

Patients

Seq Age Sex Outcome Treatment
1