ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2019-02387
- Event Type
- Malfunction
- Date Received
- April 3, 2019
- Report Date
- March 31, 2019
- Manufacturer
- ANIMAS LLC
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
DEVICE EVALUATION:IN QUARTER 2 OF 2019, THERE WERE 2 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.
OF THE 20 ALLEGATIONS OF A MALFUNCTION, 9 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 3 WERE FOUND TO BE MALFUNCTIONING DUE TO CRACKED OR TORN TRANSCEIVER FLEX. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.
THIS REPORT SUMMARIZES <NOE> 20</NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 4-70 YEARS OF AGE AND BETWEEN 100 AND 247 POUNDS. OF THE REPORTED PATIENTS, 7 WERE MALE, 13 WERE FEMALE, AND 0 WERE UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273077 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |