FDA Adverse Event Malfunction Summary report: N

BD¿ BLUNT FILL NEEDLE

MDR report key: 8476310 · Received April 2, 2019

Report

Report Number
1911916-2019-00345
Event Type
Malfunction
Date Received
April 2, 2019
Date of Event
February 22, 2019
Report Date
April 29, 2019
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
GAA
UDI-DI
30382903051800
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: FIVE (5) SAMPLES WERE RECEIVED AT BD ON 17-APR-2019 FOR INVESTIGATION IN UNOPENED BLISTER PACKS. VISUAL INSPECTION WAS PERFORMED USING 10X MAGNIFICATION. THE SAMPLES ARE NOT BENT AND NO DEFECTS WERE OBSERVED. THERE WAS NO DAMAGE; THE BEVELS AND ETCH WERE GOOD. ADDITIONALLY, NO DEFECTIVE GRIND OR HOOKS WERE NOTED. BASED ON THE INVESTIGATION CONCLUSION, THE CONDITION REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED NOR COULD THIS SYMPTOM BE CORRELATED WITH A POTENTIAL CAUSE LINKED TO THE BD PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES.

Additional Manufacturer Narrative · 0

INVESTIGATION: NO PHOTOS OR SAMPLES WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION THEREFORE THE FAILURE MODE COULD NOT BE VERIFIED AND THE ROOT CAUSE COULD NOT BE DETERMINED. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE WAS BLUNT AND BROKE WITH A BD¿ BLUNT FILL NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE HAS BEEN A LOT OF PRODUCT CONCERNS LATELY REGARDING THE BLUNT FILL NEEDLE 18G VMID 305180. IN MANY INSTANCES, THE NEEDLE WAS TOO BLUNT AND IT BROKE WHILE TRYING TO PUNCTURE THE RUBBER OF A VIAL. I HAVE NOT BEEN ABLE TO IDENTIFY IF IT IS COMING FROM A SINGLE BATCH, BUT THE PRODUCT CONCERNS ARE COMING MOSTLY FROM (B)(6) AND AT THE TIME, (B)(6) MENTIONED LOT# 8173882. I HAVE ALSO CONFIRMED THAT THERE ARE MORE INCIDENTS THAN REPORTED BELOW BUT END USERS ARE SO BUSY THAT THEY MIGHT JUST CHUCK THE DEFECTIVE PRODUCT AND USE ANOTHER ONE. I THINK IT MIGHT HAVE TO DO WITH USING THE BLUNT NEEDLE PROPERLY BECAUSE IT¿S WEIRD THE INCIDENTS ARE FROM (B)(6) ONLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE WAS BLUNT AND BROKE WITH A BD¿ BLUNT FILL NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE HAS BEEN A LOT OF PRODUCT CONCERNS LATELY REGARDING THE BLUNT FILL NEEDLE 18G VMID 305180. IN MANY INSTANCES, THE NEEDLE WAS TOO BLUNT AND IT BROKE WHILE TRYING TO PUNCTURE THE RUBBER OF A VIAL. I HAVE NOT BEEN ABLE TO IDENTIFY IF IT IS COMING FROM A SINGLE BATCH, BUT THE PRODUCT CONCERNS ARE COMING MOSTLY FROM IHA AND AT THE TIME, IHA MENTIONED LOT# 8173882. I HAVE ALSO CONFIRMED THAT THERE ARE MORE INCIDENTS THAN REPORTED BELOW BUT END USERS ARE SO BUSY THAT THEY MIGHT JUST CHUCK THE DEFECTIVE PRODUCT AND USE ANOTHER ONE. I THINK IT MIGHT HAVE TO DO WITH USING THE BLUNT NEEDLE PROPERLY BECAUSE IT¿S WEIRD THE INCIDENTS ARE FROM IHA ONLY.

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEEDLE WAS BLUNT AND BROKE WITH A BD¿ BLUNT FILL NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE HAS BEEN A LOT OF PRODUCT CONCERNS LATELY REGARDING THE BLUNT FILL NEEDLE 18G VMID (B)(4). IN MANY INSTANCES, THE NEEDLE WAS TOO BLUNT AND IT BROKE WHILE TRYING TO PUNCTURE THE RUBBER OF A VIAL. I HAVE NOT BEEN ABLE TO IDENTIFY IF IT IS COMING FROM A SINGLE BATCH, BUT THE PRODUCT CONCERNS ARE COMING MOSTLY FROM IHA AND AT THE TIME, IHA MENTIONED LOT# 8173882. I HAVE ALSO CONFIRMED THAT THERE ARE MORE INCIDENTS THAN REPORTED BELOW BUT END USERS ARE SO BUSY THAT THEY MIGHT JUST CHUCK THE DEFECTIVE PRODUCT AND USE ANOTHER ONE. I THINK IT MIGHT HAVE TO DO WITH USING THE BLUNT NEEDLE PROPERLY BECAUSE IT¿S WEIRD THE INCIDENTS ARE FROM IHA ONLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
269741 BD¿ BLUNT FILL NEEDLE MANUAL SURGICAL INSTRUMENT FOR GENERAL USE GAA BD MEDICAL (BD WEST) MEDICAL SURGICAL 8173882 30382903051800

Patients

Seq Age Sex Outcome Treatment
1 Other