FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 8474287 · Received April 2, 2019

Report

Report Number
9611109-2019-00220
Event Type
Malfunction
Date Received
April 2, 2019
Date of Event
March 5, 2019
Report Date
April 2, 2019
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). HE WAS ABLE TO CONFIRM THE REPORTED ISSUE AND TRACED THE FAILURE TO LEAKING INTERNAL COMPONENTS. THE PARTS WERE REPLACED AND THE DEVICE RETURNED TO SERVICE. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. CORRECTIVE ACTIONS ARE IN PROGRESS FOR THIS ISSUE.

Description of Event or Problem · 1

LIVANOVA DEUTSCHLAND RECEIVED A REPORT OF A WATER LEAK FROM A HEATER-COOLER SYSTEM 3T DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267566 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1