FDA Adverse Event Injury Summary report: N

AORTIC TAG THORACIC ENDOPROSTHESIS (CTAG)

MDR report key: 8472563 · Received April 2, 2019

Report

Report Number
2017233-2019-00194
Event Type
Injury
Date Received
April 2, 2019
Date of Event
March 7, 2019
Report Date
May 28, 2019
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P040043
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GR
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 0

THE FOLLOWING ARTICLE WAS REVIEWED: ¿ENDOVASCULAR REPAIR OF BLUNT THORACIC AORTIC TRAUMA: IS POSTIMPLANT HYPERTENSION AN INCIDENTAL FINDING?¿ (KONSTANTINOS TIGKIROPOULOS, ET.AL, VANN VASC SURG, 2018, 50: 160-166, PUBLISHED ONLINE ON (B)(6) 2018). THE ARTICLE ANALYZED THE OUTCOME OF 23 PATIENTS THAT PRESENTED WITH BLUNT THORACIC AORTIC TRAUMA (BTAI) THAT UNDERWENT THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) BETWEEN JANUARY 2005 AND JANUARY 2016. PATIENT 9 THAT WAS TREATED WITH TWO CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES SHOWED SIGNS OF PROXIMAL COLLAPSE OF THE ENDOGRAFT ON THE SECOND POSTOPERATIVE DAY. HE WAS URGENTLY TRANSFERRED TO THE OPERATING THEATER, AND A PROXIMAL ENDOGRAFT EXTENSION WAS PERFORMED ALONG WITH APPROPRIATE BALLOON DILATATION AND AN INNOMINATE TO THE LEFT CAROTID BYPASS. THE PHYSICIAN STATED THAT THE AORTIC ARCH OF THE PATIENT HAD AN ACUTE ANGLE THAT MIGHT HAVE CONTRIBUTED TO THE COLLAPSE OF THE DEVICE. THE PATIENT TOLERATED THE PROCEDURE.

Additional Manufacturer Narrative · 1

LOT/SERIAL NUMBERS WERE NOT PROVIDED, THEREFORE, A REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE CONDUCTED.

Description of Event or Problem · 1

THE FOLLOWING ARTICLE WAS REVIEWED: ¿ENDOVASCULAR REPAIR OF BLUNT THORACIC AORTIC TRAUMA: IS POSTIMPLANT HYPERTENSION AN INCIDENTAL FINDING?¿ (KONSTANTINOS TIGKIROPOULOS, ET.AL, VANN VASC SURG, 2018, 50: 160-166, PUBLISHED ONLINE ON MARCH 7, 2018). THE ARTICLE ANALYZED THE OUTCOME OF 23 PATIENTS THAT PRESENTED WITH BLUNT THORACIC AORTIC TRAUMA (BTAI) THAT UNDERWENT THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) BETWEEN JANUARY 2005 AND JANUARY 2016. PATIENT 9 THAT WAS TREATED WITH TWO CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES SHOWED SIGNS OF PROXIMAL COLLAPSE OF THE ENDOGRAFT ON THE SECOND POSTOPERATIVE DAY. HE WAS URGENTLY TRANSFERRED TO THE OPERATING THEATER, AND A PROXIMAL ENDOGRAFT EXTENSION WAS PERFORMED ALONG WITH APPROPRIATE BALLOON DILATATION AND AN INNOMINATE TO THE LEFT CAROTID BYPASS. THE PATIENT TOLERATED THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268535 AORTIC TAG THORACIC ENDOPROSTHESIS (CTAG) SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES

Patients

Seq Age Sex Outcome Treatment
1 31 YR Required Intervention