NEEDLE 18X1-1/2 BLUNT FILL
Report
- Report Number
- 1911916-2019-00330
- Event Type
- Malfunction
- Date Received
- April 1, 2019
- Date of Event
- February 21, 2019
- Report Date
- April 30, 2019
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- GAA
- UDI-DI
- 30382903051800
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: NO PHOTOS OR SAMPLES WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION THEREFORE THE FAILURE MODE COULD NOT BE VERIFIED AND THE ROOT CAUSE COULD NOT BE DETERMINED. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH.
INVESTIGATION SUMMARY: FIVE (5) SAMPLES WERE RECEIVED AT BD ON 17-APR-2019 FOR INVESTIGATION IN UNOPENED BLISTER PACKS. VISUAL INSPECTION WAS PERFORMED USING 10X MAGNIFICATION. THE SAMPLES ARE NOT BENT AND NO DEFECTS WERE OBSERVED. THERE WAS NO DAMAGE; THE BEVELS AND ETCH WERE GOOD. ADDITIONALLY, NO DEFECTIVE GRIND OR HOOKS WERE NOTED. BASED ON THE INVESTIGATION CONCLUSION, THE CONDITION REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED NOR COULD THIS SYMPTOM BE CORRELATED WITH A POTENTIAL CAUSE LINKED TO THE BD PROCESS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES.
IT WAS REPORTED THAT THE NEEDLE 18X1-1/2 BLUNT FILL EXPERIENCED CANNULA BREAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: MATERIAL NO: 305180 BATCH NO: 8173882. IT WAS REPORTED THAT NEEDLE WAS TOO BLUNT AND BROKE WHILE TRYING TO PUNCTURE THE RUBBER VILE. PER CUSTOMER EMAIL: THERE HAS BEEN A LOT OF PRODUCT CONCERNS LATELY REGARDING THE BLUNT FILL NEEDLE 18G VMID 305180. IN MANY INSTANCES, THE NEEDLE WAS TOO BLUNT AND IT BROKE WHILE TRYING TO PUNCTURE THE RUBBER OF A VIAL. I HAVE NOT BEEN ABLE TO IDENTIFY IF IT IS COMING FROM A SINGLE BATCH, BUT THE PRODUCT CONCERNS ARE COMING MOSTLY FROM IHA AND AT THE TIME, IHA MENTIONED LOT# 8173882.
IT WAS REPORTED THAT THE NEEDLE 18X1-1/2 BLUNT FILL EXPERIENCED CANNULA BREAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: MATERIAL NO: 305180 BATCH NO: 8173882 IT WAS REPORTED THAT NEEDLE WAS TOO BLUNT AND BROKE WHILE TRYING TO PUNCTURE THE RUBBER VILE. PER CUSTOMER EMAIL: THERE HAS BEEN A LOT OF PRODUCT CONCERNS LATELY REGARDING THE BLUNT FILL NEEDLE 18G VMID 305180. IN MANY INSTANCES, THE NEEDLE WAS TOO BLUNT AND IT BROKE WHILE TRYING TO PUNCTURE THE RUBBER OF A VIAL. I HAVE NOT BEEN ABLE TO IDENTIFY IF IT IS COMING FROM A SINGLE BATCH, BUT THE PRODUCT CONCERNS ARE COMING MOSTLY FROM IHA AND AT THE TIME, IHA MENTIONED LOT# 8173882.
DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE NEEDLE 18X1-1/2 BLUNT FILL EXPERIENCED CANNULA BREAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: MATERIAL NO: 305180, BATCH NO: 8173882. IT WAS REPORTED THAT NEEDLE WAS TOO BLUNT AND BROKE WHILE TRYING TO PUNCTURE THE RUBBER VILE. PER CUSTOMER EMAIL: THERE HAS BEEN A LOT OF PRODUCT CONCERNS LATELY REGARDING THE BLUNT FILL NEEDLE 18G VMID 305180. IN MANY INSTANCES, THE NEEDLE WAS TOO BLUNT AND IT BROKE WHILE TRYING TO PUNCTURE THE RUBBER OF A VIAL. I HAVE NOT BEEN ABLE TO IDENTIFY IF IT IS COMING FROM A SINGLE BATCH, BUT THE PRODUCT CONCERNS ARE COMING MOSTLY FROM (B)(6) AND AT THE TIME, (B)(6) MENTIONED LOT# 8173882.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 265111 | NEEDLE 18X1-1/2 BLUNT FILL | MANUAL SURGICAL INSTRUMENT FOR GENERAL USE | GAA | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 8173882 | 30382903051800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |