TECNIS 1 MULTIFOCAL
Report
- Report Number
- 9614546-2019-00283
- Event Type
- Injury
- Date Received
- April 1, 2019
- Report Date
- November 10, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- MFK
- UDI-DI
- 05050474560468
- PMA / PMN Number
- P980040
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.
DEVICE AVAILABLE FOR EVALUATION: YES. RETURNED TO MANUFACTURER ON: 4/10/2019. DEVICE RETURNED TO MANUFACTURER: YES. DEVICE EVALUATION: THE PRODUCT WAS RETURNED TO THE MANUFACTURING SITE FOR EVALUATION. THE PRODUCT WAS RECEIVED IN A ZIP LOCK BAG. VISUAL INSPECTION WITH THE UNAIDED EYE SHOWS THE PRODUCT IS CUT IN PIECES, MOST PROBABLY TO MAKE EXPLANT POSSIBLE. CONSIDERING THE CONDITION OF THE LENS ADDITIONAL ANALYSIS IS NOT POSSIBLE. THE COMPLAINT CANNOT BE CONFIRMED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE INTRAOCULAR LENS WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. A SEARCH ON COMPLAINTS REVEALED THAT NO SIMILAR COMPLAINTS WERE RECEIVED FOR THIS PRODUCTION ORDER NUMBER. LABELING REVIEW: THE DIRECTIONS FOR USE (DFU) WAS REVIEWED. THE DFU ADEQUATELY PROVIDES INSTRUCTIONS AND PRECAUTIONS ALONG WITH WARNINGS FOR THE PROPER USE AND HANDLING OF THE DEVICE. AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY AND THE REPORTED ISSUE COULD NOT BE VERIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
DATE OF EVENT: EXACT DATE UNKNOWN, NOT PROVIDED, BUT BEST ESTIMATE IS BETWEEN 1/22/2019 - 3/12/2019. (B)(4). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS EXPLANTED DUE TO AN INCORRECT LENS POWER; PATIENT WAS UNHAPPY. THE LENS WAS REPLACED WITH THE SAME MODEL LENS BUT LOWER DIOPTER. THE PATIENT REPORTED DOING FINE POST EXCHANGE. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262484 | TECNIS 1 MULTIFOCAL | MULTIFOCAL IOLS | MFK | JOHNSON & JOHNSON SURGICAL VISION, INC. | ZLB00 | 05050474560468 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |