ULTRASAFE X100L PR PLUM SSL NVS STEIN
Report
- Report Number
- 3001741852-2019-00006
- Event Type
- Malfunction
- Date Received
- March 28, 2019
- Date of Event
- March 11, 2019
- Report Date
- April 16, 2019
- Manufacturer
- BECTON DICKINSON
- Product Code
- FMF
- PMA / PMN Number
- K011369
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: BATCH RECORD REVIEW DID NOT INDICATE OF ANY INCIDENT IN RELATION TO THE PROBLEM STATEMENT. BATCH WAS RELEASED IN ACCORDANCE TO THE ACCEPTABLE CRITERIA. A SAMPLE WAS NOT RETURNED SO THE COMPLAINT COULD NOT BE CONFIRMED AND A ROOT CAUSE COULD NOT BE DETERMINED.
PER RECEIVED EMAIL, THE PMA/510(K)# HAS BEEN UPDATED. THE FOLLOWING INFORMATION HAS BEEN CORRECTED: PMA/510(K)#: K011369, K060743, K122558.
IT WAS REPORTED THAT THE ULTRASAFE X100L PR PLUM SSL NVS STEIN EXPERIENCED LEAKAGE. THE CUSTOMER STATED, "I OPENED THE PACKAGE, AND WHEN I WENT TO TAKE SYRINGE OUT. I FELT MOVEMENT AND THE PLUNGER OUT AND IT FELL ON THE FLOOR. THE THING THAT KEEP THE PLUNGER IN THERE WAS NOT IN THERE. A LOT OF THE MEDICATION IS PUSHED OUT AND THE MEDICATION GOT ON THE FLOOR AND ON ME".
IT WAS REPORTED THAT THE ULTRASAFE X100L PR PLUM SSL NVS STEIN EXPERIENCED LEAKAGE. THE CUSTOMER STATED, "I OPENED THE PACKAGE, AND WHEN I WENT TO TAKE SYRINGE OUT. I FELT MOVEMENT AND THE PLUNGER OUT AND IT FELL ON THE FLOOR. THE THING THAT KEEP THE PLUNGER IN THERE WAS NOT IN THERE. A LOT OF THE MEDICATION IS PUSHED OUT AND THE MEDICATION GOT ON THE FLOOR AND ON ME".
DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE ULTRASAFE X100L PR PLUM SSL NVS STEIN EXPERIENCED LEAKAGE. THE CUSTOMER STATED, "I OPENED THE PACKAGE, AND WHEN I WENT TO TAKE SYRINGE OUT. I FELT MOVEMENT AND THE PLUNGER OUT AND IT FELL ON THE FLOOR. THE THING THAT KEEP THE PLUNGER IN THERE WAS NOT IN THERE. A LOT OF THE MEDICATION IS PUSHED OUT AND THE MEDICATION GOT ON THE FLOOR AND ON ME".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255846 | ULTRASAFE X100L PR PLUM SSL NVS STEIN | SYRINGE | FMF | BECTON DICKINSON | 1703375 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |