FDA Adverse Event Malfunction Summary report: N

TCK1 HD CAMERA HEAD

MDR report key: 8460252 · Received March 28, 2019

Report

Report Number
1221934-2019-56700
Event Type
Malfunction
Date Received
March 28, 2019
Date of Event
December 6, 2018
Report Date
March 26, 2019
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
FWF
UDI-DI
10886705026258
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY: THE DEVICE WAS RECEIVED AT SERVICE CENTER AND EVALUATED. PER SERVICE MANUAL OPERATIONAL AND DIAGNOSTIC ANALYSIS CONFIRMED REPORTED ISSUE OF THE IMAGE CUTTING OUT. THE DEVICE WOULD INTERMITTENTLY NOT DISPLAY VIDEO OR WOULD DISPLAY POOR QUALITY VIDEO. HOWEVER,THE ROOT CAUSE FOR THE REPORTED FAILURE IS UNDETERMINED. THERE IS NO REPAIR AUTHORIZED FOR THIS DEVICE, SO THE DEVICE WILL BE PLACED INTO LONG TERM HOLD WITH NO FURTHER REPAIR ACTIVITIES TAKING PLACE. FURTHERMORE, NO NON-CONFORMANCES WERE IDENTIFIED FOR THIS PART NUMBER (242400), SERIAL NUMBER ((B)(4)) COMBINATION. REVIEW CONDUCTED PER QLIK QUERY EXECUTED ON (B)(6)2019. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE SALES REP VIA PHONE THAT DURING A SHOULDER REPAIR PROCEDURE THE CUSTOMER'S TCK1 CAMERA HEAD HAD POOR IMAGE QUALITY AND THE IMAGE WAS CUTTING OUT. THE CASE WAS COMPLETED WITH ANOTHER LIKE-DEVICE WITH NO PATIENT HARM OR DELAY. THE DEVICE IS BEING RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252188 TCK1 HD CAMERA HEAD ENDOSCOPIC VIDEO CAMERA FWF MEDOS INTERNATIONAL SARL 10886705026258

Patients

Seq Age Sex Outcome Treatment
1