FDA Adverse Event
Malfunction
Summary report: N
TAMPER-EVIDENT ORAL TIP CAP
MDR report key: 845749
·
Received April 5, 2007
Report
- Report Number
- 845749
- Event Type
- Malfunction
- Date Received
- April 5, 2007
- Date of Event
- March 20, 2007
- Report Date
- April 5, 2007
- Manufacturer
- BAXA CORPORATION
- Product Code
- KYX
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MN, US
Narratives
Description of Event or Problem · 1
RED TAMPER-EVIDENT ORAL TIP CAP WAS REMOVED FROM SYRINGE IN ORDER TO ADMINISTER MEDICATION VIA NG TUBE. WHEN CAP WAS PULLED OFF, STAFF NOTICED A PORTION OF THE CAP BROKE OFF IN THE SYRINGE. NO PATIENT HARM OCCURRED AND THE SYRINGE WAS SENT BACK TO THE PHARMACY, HOWEVER THE PRODUCT WAS NOT RETAINED. PHARMACY WAS ONLY ABLE TO DUPLICATE THE PROBLEM IF THE CAP WAS PULLED OFF AT AN ANGLE RATHER THAN STRAIGHT UP. THE MANUFACTURER HAS BEEN CONTACTED REGARDING THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAMPER-EVIDENT ORAL TIP CAP | SYRINGE CAP | KYX | BAXA CORPORATION | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2 MO |