FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER

MDR report key: 8453411 · Received March 26, 2019

Report

Report Number
1710034-2019-00331
Event Type
Malfunction
Date Received
March 26, 2019
Date of Event
March 8, 2019
Report Date
May 3, 2019
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903818341
PMA / PMN Number
K952861
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION:REVIEW OF DHR REVEALED ALL REQUIRED CHALLENGE SAMPLES, SET-UP AND IN PROCESS TESTING WAS PERFORMED IN ACCORDANCE WITH THE QUALITY PLANS. REVIEW DISCLOSED NO REJECT ACTIVITY FINDINGS THROUGHOUT THE BUILD OF THIS LOT THAT WOULD IMPACT THE OUTCOME OF THE QUALITY OF THE PRODUCT RELEVANT TO THE DEFECT. RECEIVED FOUR UNUSED IAG 20GA UNITS IN SEALED PACKAGES FROM THE CATALOG NUMBER 381834; LOT NUMBER 8173559. VISUAL/MICROSCOPIC EVALUATION: OBSERVED THERE WAS NO MECHANICAL/PHYSICAL DAMAGE TO ANY OF THE COMPONENTS (SPRING, N NEEDLE HUB, GRIPS) OR EVIDENCE OF ADHESIVE ON THE BUTTON OR HUB. FUNCTIONAL TEST (NEEDLE RETRACTION) WAS PERFORMED: PERFORMED THE HUB TWIST TEST THEN DEPRESSED THE BUTTONS. THE RETRACTIONS WERE SUCCESSFUL, NO DELAYED REACTION WAS OBSERVED. THE DEFECT NEEDLE RETRACTION FAILURE WAS NOT IDENTIFIED, REPLICATED OR CONFIRMED WITH THE RETURNED UNITS. THE RETURNED UNITS DID NOT DISPLAY ANY ADVERSE CHARACTERISTICS THAT WOULD CONTRIBUTE TO THE DEFECT THE CUSTOMER EXPERIENCED. THE ACTUAL UNIT DESCRIBED IN THE INCIDENT REPORT WAS NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER HAD A RETRACTION ISSUE. THIS OCCURRED ON 4 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED. VERBATIM: "THE SAFETY MECHANISM DELAY RETRACTED AND BLOOD SPLATTER, CAUSE BLOOD EXPOSURE TO PATIENT AND NURSE."

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER HAD A RETRACTION ISSUE. THIS OCCURRED ON 4 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED. VERBATIM: "THE SAFETY MECHANISM DELAY RETRACTED AND BLOOD SPLATTER, CAUSE BLOOD EXPOSURE TO PATIENT AND NURSE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
247635 BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 8173559 00382903818341

Patients

Seq Age Sex Outcome Treatment
1 Other