CABLE TENSIONER WITH PISTOL GRIP
Report
- Report Number
- 2939274-2019-57114
- Event Type
- Malfunction
- Date Received
- March 26, 2019
- Date of Event
- February 28, 2019
- Report Date
- February 28, 2019
- Manufacturer
- WRIGHTS LANE SYNTHES USA PRODUCTS LLC
- Product Code
- LXH
- UDI-DI
- 10886982076748
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. INVESTIGATION OF PHOTOGRAPH: NO ISSUES RELEVANT TO THE FUNCTIONAL COMPLAINT CONDITION WERE OBSERVED IN THE IMAGE PROVIDED. FADED ETCHING WAS NOTED; THE ETCHING WAS LEGIBLE. IT IS DETERMINED THAT THE REUSABLE INSTRUMENT DEVICE IS WORN FROM REPEATED USE AND SERVICING; THEREFORE, FURTHER INVESTIGATION FOR THE REPORTED COMPLAINT DEVICE IS NOT REQUIRED. CONCLUSION: THE COMPLAINT CONDITION WAS NOT CONFIRMED DURING INVESTIGATION OF THE PHOTOGRAPH. A FADED ETCH WAS NOTED. AFTER A VISUAL INSPECTION, IT IS DETERMINED THAT THE REUSABLE INSTRUMENT DEVICE IS WORN FROM REPEATED USE AND SERVICING; THEREFORE, FURTHER INVESTIGATION FOR THE REPORTED COMPLAINT DEVICE IS NOT REQUIRED. NO MANUFACTURING ISSUES WERE IDENTIFIED DURING INVESTIGATION. DURING THE INVESTIGATION, NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. THEREFORE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTIVE ACTION IS PROPOSED. DEVICE HISTORY LOT, PART NUMBER: 03.221.015, SYNTHES LOT NUMBER: P091331 , SUPPLIER LOT NUMBER: N/A, RELEASE TO WAREHOUSE DATE: 13APR2011, EXPIRATION DATE: N/A, SUPPLIER: (B)(6). NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT ON (B)(6) 2019 DURING AN OPEN REDUCTION AND INTERNAL FIXATION (ORIF) OF DISTAL FEMUR/ PERIPROSTHETIC, THE CABLE TENSIONER WITH PISTOL GRIP WAS SLIPPING AND WOULD NOT TIGHTEN DURING FINAL TENSIONING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. SURGICAL DELAY IS UNKNOWN. PATIENT STATUS IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN CABLE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) CABLE TENSIONER WITH PISTOL GRIP. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243864 | CABLE TENSIONER WITH PISTOL GRIP | MISC ORTHO SURGICAL INSTR | LXH | WRIGHTS LANE SYNTHES USA PRODUCTS LLC | 03.221.015 | P091331 | 10886982076748 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR |