FDA Adverse Event Malfunction Summary report: N

ARCHITECT C4000 ANALYZER

MDR report key: 8446722 · Received March 24, 2019

Report

Report Number
1628664-2019-00248
Event Type
Malfunction
Date Received
March 24, 2019
Date of Event
March 6, 2019
Report Date
May 16, 2019
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
JJE
UDI-DI
00380740003753
PMA / PMN Number
K980367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE SUBSEQUENT SITE VISIT, THE ABBOTT FIELD SERVICE ENGINEER (FSE) OBSERVED BUBBLES IN THE INTERNAL PROBE WASH PUMP. THE FSE REPLACED THE BELLOWS ONLY BELLOWS (ROHS) (PART NUMBER 2-89054-02) AND THE AERO C8K POPPET VALVE (LIST NUMBER 09D36-02), WHICH RESOLVED THE ISSUE. AFTER PARTS REPLACEMENT, QC WAS PERFORMED AND FOUND WITHIN ACCEPTABLE RANGES. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. A REVIEW OF THE ARCHITECT C4000, SERIAL NUMBER (B)(4), SERVICE HISTORY IDENTIFIED NO CONTRIBUTING FACTORS ON OR AROUND THE DATE OF THE COMPLAINT. THERE HAVE BEEN NO SUBSEQUENT CONTACTS FROM THE CUSTOMER REGARDING ERRATIC OR DISCREPANT GLUCOSE, TOTAL BILIRUBIN, OR QC RESULTS SINCE THE PARTS WERE REPLACED. A REVIEW OF THE ARCHITECT SYSTEM OPERATIONS MANUAL ADDRESSES OPERATIONAL PRECAUTIONS AND LIMITATIONS AS WELL AS INFORMATION REGARDING TROUBLESHOOTING ERRATIC/DISCREPANT RESULTS. THE MANUAL ALSO PROVIDES INSTRUCTION FOR THE REMOVAL, REPLACEMENT, AND VERIFICATION OF THE AERO C8K POPPET VALVE. THE ARCHITECT C4000 SYSTEM SERVICE AND SUPPORT MANUAL PROVIDES INSTRUCTION FOR THE REMOVAL, REPLACEMENT, AND VERIFICATION OF THE BELLOWS ONLY BELLOWS (ROHS). A 12-MONTH REVIEW FOR SIMILAR COMPLAINTS IDENTIFIED NO TRENDS OF THE BELLOWS ONLY BELLOWS (ROHS) OR THE AERO C8K POPPET VALVE. A REVIEW OF THE C4000 PLATFORM TRACKING AND TRENDING DATA REVEALED NO SYSTEMIC ISSUES TRENDS ASSOCIATED WITH THE DISCREPANT RESULTS DESCRIBED IN THIS COMPLAINT. BASED ON THE INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE ARCHITECT C4000, SERIAL NUMBER (B)(4), THE BELLOWS ONLY BELLOWS (ROHS) (PART NUMBER 2-89054-02), OR THE AERO C8K POPPET VALVE (LIST NUMBER 09D36-02).

Additional Manufacturer Narrative · 1

PATIENT IDENTIFIER = (B)(6). AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED BY THE CUSTOMER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSELY DEPRESSED GLUCOSE AND TOTAL BILIRUBIN RESULTS ON ONE PATIENT ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE: ON (B)(6) 2019 SID (B)(6) = GLUCOSE = 5 MG/DL 189 / 135 / 157 MG/DL; AND T. BILI = <0.1 MG/DL 0.28 MG/DL 0.36 MG/DL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
239843 ARCHITECT C4000 ANALYZER AUTOMATED CHEMISTRY ANALYZER JJE ABBOTT MANUFACTURING INC 00380740003753

Patients

Seq Age Sex Outcome Treatment
1