MENTOR SILTEX ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2019-09612
- Event Type
- Injury
- Date Received
- March 22, 2019
- Date of Event
- February 19, 2019
- Report Date
- February 25, 2019
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001812
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
ON (B)(6) 2019, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED NO FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE AND NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. DURING EVALUATION OF THE DEVICE A RENT WAS OBSERVED ON THE POSTERIOR ASPECT, MEASURING APPROXIMATELY 0.7 CM. BECAUSE THE AUTHORIZATION FOR RETURN AND EXAMINATION OF MEDICAL DEVICE FORM WAS NOT SIGNED AND/OR RETURNED TO MENTOR, MENTOR PRODUCT ANALYSIS LAB WAS PRECLUDED FROM FURTHER EVALUATING THE DEVICE TO AVOID COMPROMISE THE DEVICE INTEGRITY. NO OTHER ANOMALIES WERE OBSERVED. DEFLATION COMPLAINT WAS CONFIRMED. MENTOR PRODUCTS ARE 100% VISUALLY INSPECTED PRIOR TO RELEASE IN ADDITION TO THOROUGH IN-PROCESS TESTING DURING SEVERAL STAGES OF THE MANUFACTURING PROCESS. MENTOR PRODUCT ANALYSIS LAB CONCLUDED THAT RENTS ARE SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE. ALL THE IMPLANTS ARE 100% INSPECTED BEFORE LEAVING THE FACILITY. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED FOR THE FINISHED DEVICE NUMBER 231047, AND NO NON-CONFORMANCE'S RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. AT THIS TIME IS NOT POSSIBLE TO DETERMINE THE ORIGIN OF THE YELLOW MATERIAL. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURING RECORD EVALUATION (MRE) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6)-YEAR OLD CAUCASIAN FEMALE PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH A 375CC MENTOR SILTEX ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS, EXPERIENCED LEFT SIDE DEFLATION POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH TWO 500CC MENTOR MEMORYGEL BREAST IMPLANTS ON (B)(6) 2019.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235850 | MENTOR SILTEX ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 231047 | 00081317001812 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention |