GENERAL SPHERE INSTRUMENT SET
Report
- Report Number
- 3005180920-2019-00186
- Event Type
- Malfunction
- Date Received
- March 22, 2019
- Date of Event
- February 21, 2019
- Report Date
- March 22, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- UDI-DI
- 07630030877414
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 21 MARCH 2019: LOT 1810892: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 16-JAN-2019. EXPIRATION DATE: 2024-01-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. CONSIDERING THE (B)(4), NO OTHER SIMILAR EVENT HAS EVER BEEN REPORTED. ADDITIONAL COMPONENT THAT BROKE OFF DURING SURGERY: GMK SPHERE INSERT SET I5L-T5 REF 11.01025 (LOT 1812071): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 16-JAN-2019. EXPIRATION DATE: 2023-12-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. CONSIDERING THE (B)(4), NO OTHER SIMILAR EVENT HAS EVER BEEN REPORTED. PRELIMINARY INVESTIGATION PERFORMED BY R&D PROJECT MANAGER: ACCORDING TO WHAT REPORTED BY THE SURGEON, THE CAUSE OF THE IMPACTOR BREAKAGE WAS PROBABLY DUE TO A VERY HARD BONE; IN THIS SITUATION, THE SURGEON PROBABLY NEEDED TO APPLY EXCESSIVE FORCE DURING IMPACTIONS WITH ALSO THE RISK OF MISALIGNMENT DURING THE KEEL PREPARATION. THE HANDLE IS INTENDED TO BE IMPACTED IN ITS LONG AXIS. IMPROPER IMPACTION DIRECTION CAN OVERLOAD THE CONNECTION WITH A BENDING STRESS WITH THE CONSEQUENT RISK OF BREAKAGE OF THE HANDLE-TIP. THE POSSIBILITY TO IMMEDIATELY USE A NEW SET IN CASE OF PROBLEM, IS ONE OF THE KEY POINT OF THE EFFICIENCY INSTRUMENTATION; THE NEW DEVICE ALLOWED TO CORRECTLY REMOVE THE KEEL PUNCHER AND CONCLUDE THE SURGERY. IN CASE OF INSTRUMENT BREAKAGE OR EVEN JUST IN CASE OF ACCIDENTAL FALL OF THE INSTRUMENT, THE BEST SOLUTION IS ALWAYS TO OPEN A NEW SET INSTEAD OF TRYING TO PROCEED WITH INAPPROPRIATE TOOLS AS DONE DURING THE SURGERY.
DURING PRIMARY TKA THE TIP OF THE IMPACTOR HANDLE BROKE AND IT WAS IMPOSSIBLE TO GET THE TRIAL KEEL OUT OF THE PATIENT. AFTER SOME ATTEMPTS A METAL CHISEL WAS USED TO GET IT OUT. ALSO IN THIS CASE IT WAS IMPOSSIBLE TO REMOVE THE TRIAL KEEL, THAT BROKE OFF, TOO. THEY WERE FORCED TO OPEN A NEW GMK EFFICIENCY KIT AND USE THE NEW IMPACTOR HANDLE SUCCESSFULLY. THE BONE WAS VERY HARD DUE TO A TIBIA OSTEOTOMY CARRIED OUT SOME YEARS AGO BEFORE THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235562 | GENERAL SPHERE INSTRUMENT SET | SURGICAL INSTRUMENT FOR KNEE | LXH | MEDACTA INTERNATIONAL SA | 1810892 | 07630030877414 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |