FDA Adverse Event Injury Summary report: N

APDP BASE BP EN ES USA

MDR report key: 8443430 · Received March 22, 2019

Report

Report Number
1419937-2019-00046
Event Type
Injury
Date Received
March 22, 2019
Date of Event
March 1, 2019
Manufacturer
MEDELA LLC
Product Code
HGX
PMA / PMN Number
K031614
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER SERVICE CONDUCTED TROUBLESHOOTING WITH THE CUSTOMER, INCLUDING POWERING THE DEVICE ON WITH THE POWER SUPPLY, INSPECTING THE FACEPLATE AND BACK PLATE FOR DAMAGE; INSPECTING AND CLEANING THE DIAPHRAGM; DISASSEMBLING, INSPECTING, CLEANING AND REASSEMBLING THE KIT AND TUBING SET; AND VERIFYING BREAST SHIELD FIT. THE CUSTOMER WAS SENT A REPLACEMENT PUMP AND RETURN OF HER ORIGINAL PUMP WAS REQUESTED FOR TESTING/EVALUATION. THE CUSTOMER WAS CONTACTED BY A COMPLAINT HANDLER ON MULTIPLE OCCASIONS, INCLUDING IN WRITING, TO GET ADDITIONAL INFORMATION, WITH NO RESPONSE AS OF THE DATE OF THIS REPORT. THE DEVICE WAS RETURNED WITH THE CUSTOMER'S PARTS AND ACCESSORIES AND WAS EVALUATED ON 03/20/2019. THE DEVICE PASSED SUCTION AND CYCLE SPECIFICATIONS, THOUGH IT WAS IDENTIFIED IN THE EVALUATION THAT THE CUSTOMER'S VALVES AND CONNECTORS WERE DIRTY. REFER TO ATTACHED EVALUATION. THE CUSTOMER'S REPORT OF LOW SUCTION COULD NOT BE CONFIRMED. MEDELA IS FILING THIS REPORT, WHICH IS CONSIDERED A SERIOUS INJURY AS IT REQUIRED MEDICAL ATTENTION (MEDICATION WAS PRESCRIBED).

Description of Event or Problem · 1

ON (B)(6) 2019, THE CUSTOMER ALLEGED TO MEDELA LLC THAT HER PUMP IN STYLE BREAST PUMP WAS LOSING SUCTION. THE CUSTOMER INDICATED THAT SHE SINGLE PUMPS ONE SIDE AT A TIME AND SHE USED TO GET 3 OUNCES IN 5 MINUTES PER SIDE, BUT IT IS NOW TAKING HER 10 MINUTES. THE CUSTOMER ADDITIONALLY ALLEGED THAT SHE HAD CLOGGED DUCTS FOR WHICH SHE WAS PRESCRIBED ANTIBIOTICS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
238519 APDP BASE BP EN ES USA PUMP, BREAST, POWERED HGX MEDELA LLC 101035078 448568

Patients

Seq Age Sex Outcome Treatment
1 Other