FDA Adverse Event Injury Summary report: N

IL GEM PREMIER 3000 BLOOD GAS ANALYZER WITH IONIZED

MDR report key: 844287 · Received April 30, 2007

Report

Report Number
MW1042747
Event Type
Injury
Date Received
April 30, 2007
Date of Event
January 1, 2007
Report Date
April 30, 2007
Manufacturer
INSTRUMENTATION LABORATORY
Product Code
CHL
Product Problem
Yes
Report Source
Voluntary report
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

SINCE OUR STARTING TO USE OF THE IL GEM 3000 SYSTEM (FOR ABOUT ONE YEAR), WE HAVE NOTICED HIGH IONIZED CALCIUM VALUES WHENEVER A NEW REAGENT CARTRIDGE IS PLACED INTO SERVICE. THESE HIGH VALUES ARE NOTED WITH BOTH QUALITY CONTROL SPECIMENS AND PATIENT SPECIMENS. IF THESE HIGHER SPECIMENS ARE RUN ON THE OTHER IL GEM 3000 SYSTEM (OLDER REAGENT CARTRIDGE), THE VALUES FALL TO THE EXPECTED VALUE. THE DURATION OF THE ERRONEOUSLY HIGH VALUES IS QUITE SHORT AND PROBABLY CAN BE ASCERTAINED BY THE ANALYSIS OF QUALITY CONTROL SPECIMENS. THE SIZE OF THE POSITIVE BIAS IS ABOUT 0.08 AT CONCENTRATIONS OF 0.8 AND 1.25 MMOL/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IL GEM PREMIER 3000 BLOOD GAS ANALYZER WITH IONIZED * CHL INSTRUMENTATION LABORATORY * *
2 IL GEM PREMIER 3000 BLOOD GAS ANALYZER WITH IONIZED * CHL * * *

Patients

Seq Age Sex Outcome Treatment
1 * Disability