IL GEM PREMIER 3000 BLOOD GAS ANALYZER WITH IONIZED
MDR report key: 844287
·
Received April 30, 2007
Report
- Report Number
- MW1042747
- Event Type
- Injury
- Date Received
- April 30, 2007
- Date of Event
- January 1, 2007
- Report Date
- April 30, 2007
- Manufacturer
- INSTRUMENTATION LABORATORY
- Product Code
- CHL
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
SINCE OUR STARTING TO USE OF THE IL GEM 3000 SYSTEM (FOR ABOUT ONE YEAR), WE HAVE NOTICED HIGH IONIZED CALCIUM VALUES WHENEVER A NEW REAGENT CARTRIDGE IS PLACED INTO SERVICE. THESE HIGH VALUES ARE NOTED WITH BOTH QUALITY CONTROL SPECIMENS AND PATIENT SPECIMENS. IF THESE HIGHER SPECIMENS ARE RUN ON THE OTHER IL GEM 3000 SYSTEM (OLDER REAGENT CARTRIDGE), THE VALUES FALL TO THE EXPECTED VALUE. THE DURATION OF THE ERRONEOUSLY HIGH VALUES IS QUITE SHORT AND PROBABLY CAN BE ASCERTAINED BY THE ANALYSIS OF QUALITY CONTROL SPECIMENS. THE SIZE OF THE POSITIVE BIAS IS ABOUT 0.08 AT CONCENTRATIONS OF 0.8 AND 1.25 MMOL/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IL GEM PREMIER 3000 BLOOD GAS ANALYZER WITH IONIZED | * | CHL | INSTRUMENTATION LABORATORY | * | * | |
| 2 | IL GEM PREMIER 3000 BLOOD GAS ANALYZER WITH IONIZED | * | CHL | * | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Disability |